J Bone Joint Surg Br. 1988 May;70(3):397-403. doi: 10.1302/0301-620X.70B3.3372559.
This study aimed to compare the holding strength of various commercially-available anchorage devices for artificial ligaments, so that surgeons might make a reasoned choice. Tensile tests to failure were performed on screws, bollards, toggles and staples which had been implanted into cadaveric bones. The holding strength of all devices correlated significantly with the local thickness of cortical bone, so it is recommended that anchorages should be placed away from the joint line, into diaphyseal bone if possible. A new trans-cortical grommet was developed which, when used around an AO screw, had significantly greater holding strength than the other devices.