Low-Voltage Electron-Probe Microanalysis of Uranium.

作者信息

Matthews Mike B, Kearns Stuart L, Buse Ben

机构信息

AWE, Aldermaston, ReadingRG7 4PR, UK.

School of Earth Sciences, University of Bristol, Wills Memorial Building, Queens Road, CliftonBS8 1RJ, UK.

出版信息

Microsc Microanal. 2021 Mar 18:1-18. doi: 10.1017/S1431927621000192.

Abstract

Electron-probe microanalysis of uranium and uranium alloys poses several problems, such as rapid oxidation, large poorly constrained correction factors, and a large number of characteristic x-ray lines. We show that U-metal can grow 10 nm of oxide within ~20 s of air exposure, increasing to 15–20 nm within a few minutes, which can produce a 30% quantification error at 5 kV. A 15 nm carbon coating on the UO2 reference material also produces an ~30% quantification error of the uncoated but surface oxidized U sample at 5 kV. Correcting for both the coating and oxide improved the analysis accuracy to better than ±1% down to 7 kV and ~2% at 5 kV, but the error increases strongly below this. The measurement of C in U identified a previously unreported U N6–O4 line interference on the C Kα peak, which can produce over 1% error in the analysis total. Oxide stoichiometry was demonstrated to have only a small impact on quantification. The measurement of the O Kα and U Mα mass absorption coefficients in U as 9,528 and 798 cm2/g, respectively, shows good agreement with recently published values and also produces small differences in a quantification error.