Suppr超能文献

新冠康复期血浆样本中自动化化学发光法与酶联免疫吸附测定(ELISA)抗体滴度的相关性:快速、经济高效的半定量诊断方法的开发

Correlation of Automated Chemiluminescent Method with Enzyme-Linked Immunosorbent Assay (ELISA) Antibody Titers in Convalescent COVID-19 Plasma Samples: Development of Rapid, Cost-Effective Semi-Quantitative Diagnostic Methods.

作者信息

Mendoza Rachelle, Silver Michael, Zuretti Alejandro R, Christian Manan, Das Ballabh, Norin Allen J, Borgen Patrick, Libien Jenny, Bluth Martin H

机构信息

Department of Pathology, SUNY Downstate Health Sciences University, Brooklyn, NY, USA.

School of Public Health, SUNY Downstate Health Sciences University, Brooklyn, NY, USA.

出版信息

J Blood Med. 2021 Mar 17;12:157-164. doi: 10.2147/JBM.S296730. eCollection 2021.

Abstract

BACKGROUND

We investigated the utility of an automated chemiluminescent SARS-CoV-2 IgG antibody assay platform in quantifying the amount of binding antibodies present in donated convalescent plasma.

METHODS

A total of 179 convalescent plasma units were analyzed for the presence of SARS-CoV-2 IgG antibodies using the Beckman-Coulter chemiluminescent immunoassay (CLIA) platform. The equipment-derived numerical values (S/Co ratio) were recorded. Aliquots from the same units were subjected to enzyme-linked immunosorbent assay (ELISA) that detects IgG antibodies against the receptor-binding domain (RBD) of the SARS-CoV-2 S1 protein. The relationship between ELISA titers and CLIA S/Co values was analyzed using linear regression and receiver operating characteristics (ROC) curve.

RESULTS

Twenty-one samples (11.7%) had S/Co values of less than 1.0 and were deemed negative for antibodies and convalescent plasma had S/Co values between >1.0 and 5.0 (70/179, 39.1%). Fifteen units (8.4%) had negative ELISA titer. The majority of the units (95/179. 53.1%) had titers ≥1:1024. The sensitivities of ELISA to CLIA were comparable (90.5% vs 88.3%, respectively; p=0.18). There was positive linear correlation between CLIA S/Co values and ELISA IgG titer (Rho = 0.75; Spearman's rank = 0.82, p-value = <0.0001). The agreement between the two methods was fair, with a κ index of 0.2741. Using the ROC analysis, we identified a CLIA S/Co cutoff value of 8.2, which gives a sensitivity of 90% and a specificity of 82% in predicting a titer dilution of ≥1:1024.

CONCLUSION

The utility of automated antibody detection systems can be extended from simply a screening method to a semi-quantitative and quantitative functional antibody analysis. CLIA S/Co values can be used to reliably estimate the ELISA antibody titer. Incorporation of chemiluminescent-based methods can provide rapid, cost-effective means of identifying anti-SARS-CoV-2 antibody titers in donated plasma for use in the treatment of COVID-19 infection.

摘要

背景

我们研究了一种自动化化学发光SARS-CoV-2 IgG抗体检测平台在定量检测康复期血浆中结合抗体量方面的实用性。

方法

使用贝克曼库尔特化学发光免疫分析(CLIA)平台对总共179个康复期血浆样本进行SARS-CoV-2 IgG抗体检测。记录仪器得出的数值(S/Co比值)。对同一血浆样本的等分试样进行酶联免疫吸附测定(ELISA),以检测针对SARS-CoV-2 S1蛋白受体结合域(RBD)的IgG抗体。使用线性回归和受试者工作特征(ROC)曲线分析ELISA滴度与CLIA S/Co值之间的关系。

结果

21个样本(11.7%)的S/Co值小于1.0,被判定为抗体阴性,康复期血浆的S/Co值在>1.0至5.0之间(70/179,39.1%)。15个样本(8.4%)的ELISA滴度为阴性。大多数样本(95/179,53.1%)的滴度≥1:1024。ELISA与CLIA的敏感性相当(分别为90.5%和88.3%;p=0.18)。CLIA S/Co值与ELISA IgG滴度之间存在正线性相关性(Rho = 0.75;Spearman秩相关系数 = 0.82,p值 = <0.0001)。两种方法之间的一致性一般,κ指数为0.2741。通过ROC分析,我们确定CLIA S/Co临界值为8.2,在预测滴度稀释≥1:1024时,其敏感性为90%,特异性为82%。

结论

自动化抗体检测系统的用途可以从简单的筛查方法扩展到半定量和定量功能抗体分析。CLIA S/Co值可用于可靠地估计ELISA抗体滴度。采用基于化学发光的方法可以提供快速、经济有效的手段来确定捐赠血浆中抗SARS-CoV-2抗体滴度,用于治疗COVID-19感染。