进行性低氧血症期间的耳部血氧测定法。

Ear oximetry during progressive hypoxia.

作者信息

Gislason T, Sandhagen B, Daskalopouloul E

机构信息

Department of Lung Medicine, Uppsala University, Sweden.

出版信息

Ups J Med Sci. 1988;93(1):45-51. doi: 10.1517/03009734000000036.

Abstract

The BIOX III Pulse oximeter for measuring arterial oxygen saturation (SaO2) was compared during rest and under progressively hypoxic conditions, with SaO2 values based on arterial blood samples. The measurements were performed in 16 subjects undergoing tests of ventilatory response to hypoxia, by a rebreathing method. For each individual subject, there was a linear response relationship (r = 0.99), while for all 126 comparative values the regression equation was: y = 0.83 X + 14.7 (r = 0.98). The observed ventilatory response was lower when the calculations were based on oximeter readings. We conclude that the oximeter has acceptable correlation between the BIOX III and SaO2 measurements for clinical use, especially when SaO2 is above 70%.

摘要

采用再呼吸法,对16名接受低氧通气反应测试的受试者,在静息状态和逐渐缺氧的条件下,使用用于测量动脉血氧饱和度(SaO2)的BIOX III脉搏血氧仪,并将其测量值与基于动脉血样本的SaO2值进行比较。对于每个个体受试者,存在线性反应关系(r = 0.99),而对于所有126个比较值,回归方程为:y = 0.83X + 14.7(r = 0.98)。当基于血氧仪读数进行计算时,观察到的通气反应较低。我们得出结论,该血氧仪在BIOX III测量值与SaO2测量值之间具有可接受的相关性,适用于临床使用,尤其是当SaO2高于70%时。