Sidi A, Rush W, Gravenstein N, Ruiz B, Paulus D A, Davis R F
Department of Anesthesiology, College of Medicine, University of Florida, Gainesville 32610-0254.
J Clin Monit. 1987 Oct;3(4):257-62. doi: 10.1007/BF03337380.
The accuracy of two commercially available pulse oximeters (the Ohmeda Biox 3700, software version "J," and the Nellcor N-100) in detecting low levels of arterial hemoglobin oxygen saturation (SaO2) was evaluated in 10 dogs in which hypoxia was induced by stopping the fresh gas flow into the anesthesia machine circle system. Measurements made in vivo with the pulse oximeters, with detectors placed on the tongue, were compared with measurements made in vitro using an IL 282 CO-Oximeter as SaO2 decreased toward zero. Measurements from the two oximeters correlated poorly over the range from 0 to 100% SaO2 (r = 0.69). In this range, the correlation between Nellcor N-100 measurements and those of the CO-Oximeter had an r of 0.82, a regression line slope of 0.82, and a y intercept of 14.8; the correlation between the Ohmeda Biox 3700 and the CO-Oximeter had an r of 0.83, a regression line slope of 0.66, and a y intercept of 32.7. The correlation with the CO-Oximeter was similar for both the Ohmeda and the Nellcor pulse oximeters at an SaO2 of 80% or more. However, when SaO2 was less than 80%, measurements by pulse oximetry correlated less well with CO-Oximeter measurements (r = 0.62, slope = 0.64, and y intercept = 21.0 for Nellcor; r = 0.71, slope = 0.67, and y intercept = 32.4 for Ohmeda). When SaO2 was less than 60%, both oximeters inaccurately indicated the co-oximetry values (r = 0.36 and y intercept = 26.1 for the Nellcor; r = 0.48 and y intercept = 33.2 for the Ohmeda).(ABSTRACT TRUNCATED AT 250 WORDS)
在10只狗中评估了两种市售脉搏血氧仪(Ohmeda Biox 3700,软件版本“J”,以及Nellcor N - 100)检测低水平动脉血红蛋白氧饱和度(SaO2)的准确性,这些狗通过停止向麻醉机环路系统输送新鲜气体来诱导缺氧。随着SaO2降至零,将置于舌头上的探测器用脉搏血氧仪在体内进行的测量与使用IL 282 一氧化碳血氧仪在体外进行的测量进行比较。在0至100% SaO2范围内,两种血氧仪的测量结果相关性较差(r = 0.69)。在此范围内,Nellcor N - 100测量值与一氧化碳血氧仪测量值之间的相关性r为0.82,回归线斜率为0.82,y轴截距为14.8;Ohmeda Biox 3700与一氧化碳血氧仪之间的相关性r为0.83,回归线斜率为0.66,y轴截距为32.7。当SaO2为80%或更高时,Ohmeda和Nellcor脉搏血氧仪与一氧化碳血氧仪的相关性相似。然而当SaO2低于80%时,脉搏血氧测量值与一氧化碳血氧仪测量值的相关性较差(Nellcor的r = 0.62,斜率 = 0.64,y轴截距 = 21.0;Ohmeda的r = 0.71,斜率 = 0.67,y轴截距 = 32.4)。当SaO2低于60%时,两种血氧仪均不准确地显示一氧化碳血氧测定值(Nellcor的r = 0.36,y轴截距 = 26.1;Ohmeda的r = 0.48,y轴截距 = 33.2)。(摘要截短于250字)