To compare the measurement of central thickness and depth of involvement of opacity-bearing corneas at different intensities (mild, moderate, and dense) using different instruments: Visante and Optovue OCTs, ultrasound biomicroscopy (UBM), and ultrasound pachymetry (central thickness).
Sample of 102 eyes: 70 eyes (68.63 %) with corneal opacity; 32 eyes (31.37%) with normal corneas. Corneal opacity grading included mild (28 eyes, 40.00 %), moderate (27 eyes, 37.57 %), and dense (15 eyes, 21.43 %). Opacity intensity was graded and documented. Central corneal thickness was determined using Optovue and Visante OCTs, ultrasound pachymetry, and UBM VuMax (50 MHz), and depth of corneal opacity, using Optovue and Visante OCTs and UBM.
Total corneal thickness of the control group showed differences with a small correction factor using Optovue OCT (534.03 ± 39.88 μm), Visante OCT (523.72±38.70 μm), and ultrasound pachymetry (529.84 ± 39.76 μm), and were higher when compared to UBM (492.06 ± 37.93 μm). In mild opacity, depth measurements were the same with OCT Optovue and Visante and higher than those by UBM. In moderate opacities, there were no differences in measurements by different instruments. In dense opacities, OCT Optovue and Visante measurements were the same and higher than those by UBM.
Central corneal thickness measurements were considered higher in corneas with opacity than in normal corneas. We observed that the greatest thicknesses were measured in corneas with dense opacities. There was a difference between the measurements taken by different instruments, both in central thickness and in opacity depth in cases of mild and dense opacity, and no difference in moderate opacity.
使用不同仪器(Visante 和 Optovue OCT、超声生物显微镜(UBM)和超声角膜测厚仪(中央厚度))比较不同强度(轻度、中度和重度)混浊角膜的中央厚度和混浊深度的测量值。
共纳入 102 只眼,其中 70 只眼(68.63%)有角膜混浊,32 只眼(31.37%)为正常角膜。角膜混浊分级包括轻度(28 只眼,40.00%)、中度(27 只眼,37.57%)和重度(15 只眼,21.43%)。混浊强度分级并记录。采用 Optovue 和 Visante OCT、超声角膜测厚仪和 UBM VuMax(50MHz)测量中央角膜厚度,采用 Optovue 和 Visante OCT 和 UBM 测量角膜混浊深度。
对照组总角膜厚度存在差异,Optovue OCT(534.03±39.88μm)、Visante OCT(523.72±38.70μm)和超声角膜测厚仪(529.84±39.76μm)校正小误差因子后,均高于 UBM(492.06±37.93μm)。在轻度混浊中,OCT Optovue 和 Visante 的深度测量值相同,高于 UBM。在中度混浊中,不同仪器的测量值无差异。在重度混浊中,OCT Optovue 和 Visante 的测量值相同,高于 UBM。
与正常角膜相比,混浊角膜的中央角膜厚度测量值更高。我们观察到,在重度混浊的角膜中测量到的厚度最大。在轻度和重度混浊的情况下,不同仪器的测量值(包括中央厚度和混浊深度)存在差异,而在中度混浊的情况下,测量值无差异。