The influence of migration of the exeter V40 stem on patient reported outcome measures: a 2-year follow-up of 112 total hip arthroplasties using radiostereometric analysis.

机构信息

Department of Orthopaedic Surgery, Aalborg University Hospital, Farsoe, Denmark.

Department of Orthopaedic Surgery, School of Medicine, Aalborg University, Aalborg, Denmark.

出版信息

Eur J Orthop Surg Traumatol. 2022 Jan;32(1):167-174. doi: 10.1007/s00590-021-02937-x. Epub 2021 Mar 25.

Abstract

PURPOSE

The cemented Exeter V40 stem is known to migrate distally. Several previous studies have reported on the extent of stem migration and its influence on clinical outcome. However, no studies have investigated the influence of stem migration on Patient Reported Outcome Measures (PROM).

METHODS

One hundred and twelve total hip arthroplasties (THA) were included in a 2-year follow-up using Radiostereometric Analysis (RSA). Patients were evaluated using the Oxford Hip Score (OHS) and EQ-5D-3L PROMs. The purpose of this study was to assess the influence of stem migration, measured by Maximum Total Point Motion (MTPM), on the 2-year postoperative score (OHS and EQ-5D). Furthermore, the influence of pre-operative PROM, age, gender, acetabular component and BMI was associated with the 2-year postoperative OHS and EQ-5D scores.

RESULTS

MTPM was a non-significant predictor of the 2-year postoperative OHS (regression coefficient (B) = - 2.38 (CI - 5.44; .69)) and of the 2-year postoperative EQ-5D (B = - .01 (CI - .04; .02)). The only significant predictor of the 2-year postoperative OHS and 2-year postoperative EQ-5D was gender (B = 8.71 (CI 3.52; 13.90)) and (B = .13 (CI .07; .18)), respectively.

CONCLUSION

Stem migration did not significantly influence PROMs at 2 years post-operatively. Using a patient-focused approach, our results seem to corroborate results reported by previous studies, showing that slow migration of the Exeter V40 stem does not seem to influence the clinical outcome.

摘要

目的

已证实骨水泥型 Exeter V40 柄会发生向远端迁移。先前有几项研究报道了柄的迁移程度及其对临床结果的影响。然而,目前尚无研究探讨柄迁移对患者报告的结果测量(PROM)的影响。

方法

本研究共纳入了 112 例行全髋关节置换术(THA)患者,随访时间为 2 年,采用放射学体层摄影术(RSA)进行测量。使用牛津髋关节评分(OHS)和 EQ-5D-3L PROM 对患者进行评估。本研究旨在评估最大总点移动(MTPM)测量的柄迁移对术后 2 年评分(OHS 和 EQ-5D)的影响。此外,还分析了术前 PROM、年龄、性别、髋臼部件和 BMI 与术后 2 年 OHS 和 EQ-5D 评分的相关性。

结果

MTPM 是术后 2 年 OHS(回归系数(B)=-2.38(CI-5.44;.69))和术后 2 年 EQ-5D(B=-.01(CI-.04;.02))的非显著预测因子。术后 2 年 OHS 和 EQ-5D 的唯一显著预测因子是性别(B=8.71(CI3.52;13.90))和(B=.13(CI.07;.18))。

结论

柄迁移在术后 2 年内对 PROM 无显著影响。采用以患者为中心的方法,我们的结果似乎与先前研究的结果相符,表明 Exeter V40 柄的缓慢迁移似乎不会影响临床结果。