Am J Vet Res. 2021 Apr;82(4):268-273. doi: 10.2460/ajvr.82.4.268.
To evaluate the repeatability and accuracy of fingertip pulse oximeters (FPO) for measurement of hemoglobin oxygen saturation in arterial blood and pulse rate (PR) in anesthetized dogs breathing 100% O.
29 healthy client-owned anesthetized dogs undergoing various surgical procedures.
In randomized order, each of 7 FPOs or a reference pulse oximeter (PO) was applied to the tongue of each intubated anesthetized dog breathing 100% O. Duplicate measurements of oxygen saturation (Spo) and PR were obtained within 60 seconds of applying an FPO or PO. A nonparametric version of Bland-Altman analysis was used. Coefficient of repeatability was the interval between the 5th and 95th percentiles of the differences between duplicate measurements. Bias was the median difference, and the limits of agreement were the 5th and 95th percentiles of the differences between each FPO and the PO. Acceptable values for the coefficient of repeatability of Spo were ≤ 6%. Agreements were accepted if the limits of agreement had an absolute difference of ≤ ± 3% in Spo and relative difference of ≤ ± 10% in PR.
Coefficient of repeatability for Spo was acceptable for 5 FPOs, but the limits of agreement for Spo were unacceptable for all FPOs. The limits of agreement for PR were acceptable for 2 FPOs.
Results suggested that some FPOs may be suitable for accurately monitoring PRs of healthy anesthetized dogs breathing 100% O, but mild underestimation of Spo was common.
评估指夹式脉搏血氧仪(FPO)在 100%氧气通气下测量麻醉犬动脉血氧饱和度(SpO2)和脉搏率(PR)的可重复性和准确性。
29 只接受各种手术的健康有主麻醉犬。
以随机顺序,每个 FPO 或参考脉搏血氧仪(PO)应用于每个接受 100%氧气通气的麻醉犬的舌头上。在应用 FPO 或 PO 后 60 秒内,对 SpO2 和 PR 进行重复测量。使用非参数版 Bland-Altman 分析。重复性系数是重复测量差值的第 5 个和第 95 个百分位数之间的间隔。偏差是中位数差值,一致性界限是每个 FPO 与 PO 之间差值的第 5 个和第 95 个百分位数。Spo2 重复性系数的可接受值为≤6%。如果一致性界限在 SpO2 上的绝对差值≤±3%,在 PR 上的相对差值≤±10%,则认为协议可以接受。
5 个 FPO 的 Spo2 重复性系数可接受,但所有 FPO 的 Spo2 一致性界限不可接受。2 个 FPO 的 PR 一致性界限可接受。
结果表明,一些 FPO 可能适用于准确监测 100%氧气通气下健康麻醉犬的 PR,但 SpO2 轻度低估是常见的。