Gould L, Betzu R, Judge D, Lee J, Taddeo M
Methodist Hospital, Brooklyn, New York.
Angiology. 1988 May;39(5):409-12. doi: 10.1177/000331978803900501.
The porous-tip electrode has led to a significant advancement in pacing system technology. Experience with 38 patients with unipolar tined CPI porous endocardial electrodes, is reported. All patients had programmable pulse generators, CPI model number 531. The voltage in this pulse generator can be reduced from 5 to 2.5 volts. The pulse width threshold (PWT) at implantation was 0.06 ms at 5 volts in all 38 patients, and the PWT at implantation varied between 0.06 and 0.16 ms at 2.5 volts, with a mean PWT of 0.092 +/- 0.026 ms. The 38 patients were restudied six months later. The mean PWT measured at 5 volts was 0.076 +/- 0.018 ms. The PWT measured at 2.5 volts varied between 0.08 and 0.30 ms with a mean PWT of 0.178 +/- 0.067 ms. Twenty-three patients were restudied one year after lead implantation. The PWT measured at 2.5 volts varied between 0.08 and 0.30 ms with a mean PWT of 0.156 +/- 0.061 ms. All of the pulse generators were reprogrammed to 2.5 volts. This permitted a substantial prolongation of the pulse generator life (three to five years) with an improvement in the cost effectiveness of the pulse generator.
多孔尖端电极使起搏系统技术取得了重大进展。本文报告了38例使用单极带倒刺CPI多孔心内膜电极的患者的经验。所有患者均使用可编程脉冲发生器,型号为CPI 531。该脉冲发生器的电压可从5伏降至2.5伏。在植入时,所有38例患者在5伏电压下的脉冲宽度阈值(PWT)为0.06毫秒,在2.5伏电压下植入时的PWT在0.06至0.16毫秒之间变化,平均PWT为0.092±0.026毫秒。六个月后对这38例患者进行了再次研究。在5伏电压下测得的平均PWT为0.076±0.018毫秒。在2.5伏电压下测得的PWT在0.08至0.30毫秒之间变化,平均PWT为0.178±0.067毫秒。在导线植入一年后对23例患者进行了再次研究。在2.5伏电压下测得的PWT在0.08至0.30毫秒之间变化,平均PWT为0.156±0.061毫秒。所有脉冲发生器均重新编程为2.5伏。这使得脉冲发生器的使用寿命大幅延长(三至五年),同时提高了脉冲发生器的性价比。