Several pre-analytical factors can affect the measurement of intact Parathyroid Hormone (IPTH). In this study, we have investigated the effects of using different types of tubes, time elapsed before separation, and storage conditions over time on the measured values of IPTH.
Blood samples from 30 subjects were collected into plain, SST, and EDTA tubes. All serum and plasma were separated immediately (first set) and after 2 hrs delay (second set). The first set of samples were aliquoted and stored at RT (25°C), at fridge (4°C), and freezer (-20°C). IPTH was measured in all the stored aliquots at 2,4, and 8 days after collection using Architect analyzer.
Paired T test and ANOVA repeated measures showed no significant difference between IPTH levels in all tubes. The second set of serum and plasma were significantly lower (3.8% and 7.4%, p < 0.001, respectively) when compared to samples measured initially. Serum samples stored at RT were significantly lower (by 45%,59%, and 77%) on days 2,4, and 8 when compared to the initial time (p < 0.001 in all cases). Plasma samples stored at RT, were significantly lower on day 8 after collection, by 30.8% (p < 0.001). These differences would be clinically important.
Plasma IPTH can be stored at RT for up to four days. Both plasma and serum IPTH are not affected by a delay in the separation of up to two h and they can be stored for up to 8 days in a fridge or freezer without any clinically significant changes in their values.
几个分析前因素会影响完整甲状旁腺激素(IPTH)的测量。在这项研究中,我们研究了使用不同类型的采血管、分离前的时间延迟以及随时间的储存条件对 IPTH 测量值的影响。
从 30 名受试者中采集血样到普通管、SST 管和 EDTA 管中。所有血清和血浆均立即(第一组)和 2 小时后(第二组)分离。第一组样本等分并在 RT(25°C)、冰箱(4°C)和冷冻室(-20°C)下储存。在收集后 2、4 和 8 天,使用 Architect 分析仪测量所有储存等分样本中的 IPTH。
配对 T 检验和方差分析重复测量显示,所有采血管中的 IPTH 水平之间没有显著差异。与初始测量的样本相比,第二组血清和血浆显著降低(分别为 3.8%和 7.4%,p<0.001)。与初始时间相比,在第 2、4 和 8 天,在 RT 下储存的血清样本分别显著降低(所有情况下均 p<0.001),降低幅度为 45%、59%和 77%。在 RT 下储存的血浆样本在收集后第 8 天显著降低,降低幅度为 30.8%(p<0.001)。这些差异具有临床意义。
血浆 IPTH 可以在 RT 下储存长达 4 天。血浆和血清 IPTH 不受分离延迟长达 2 小时的影响,并且可以在冰箱或冷冻室中储存长达 8 天,而其值没有任何临床显著变化。