Chen Jian, Cheng Hu, Zhou Xuefeng, Yan Xiaozhi, Wang Lingfei, Zhao Yusheng, Wang Shanmin
Department of Physics, Southern University of Science and Technology, Shenzhen, Guangdong 518055, China.
Rev Sci Instrum. 2021 Mar 1;92(3):033905. doi: 10.1063/5.0040847.
Pressure calibration for most diamond-anvil cell (DAC) experiments is mainly based on the ruby scale, which is key to implementing this powerful tool for high-pressure study. However, the ruby scale can often hardly be used for programmably controlled DAC devices, especially the piezoelectric-driving cells, where a continuous pressure calibration is required. In this work, we present an effective pressure gauge for DACs made of Manganin metal based on the four-probe resistivity measurements. Pressure dependence of its resistivity is well established and shows excellent linear relations in the 0-30 GPa pressure range with a slope of 23.4 (9) GPa for the first-cycle compression, in contrast to that of multiple-cycle compression and decompression having a nearly identical slope of 33.7 (4) GPa likely due to the strain effect. In addition, the such-established Manganin scale can be used for continuously monitoring the cell pressure of piezoelectric-driving DACs, and the reliability of this method is also verified by the fixed-point method with a Bi pressure standard. Realization of continuous pressure calibration for programmably controlled DACs would offer many opportunities for the study of dynamics, kinetics, and critical behaviors of pressure-induced phase transitions.