Perkins Thomas J, Kop Alan M, Whitewood Colin, Pabbruwe Moreica B
University of Western Australia, Crawley, WA, Australia.
Department of Medical Engineering and Physics, Centre for Implant Technology and Retrieval Analysis, Royal Perth Hospital, Perth, Australia.
Hip Int. 2023 Jan;33(1):28-33. doi: 10.1177/11207000211008459. Epub 2021 Apr 9.
Although the Pinnacle Acetabular Hip System (DePuy Synthes) has demonstrated excellent survivorship results since it was first introduced in 2003, there have been a growing number of cases indicating that Pinnacle liners may be subject to a higher-than-expected rate of early dissociation failure. Between 2006 and 2020, our Centre received 212 retrieved Pinnacle liners from Western Australian hospitals. Of these, 26 were removed due to liner dissociation.
To better understand the frequency and cause of this complication we assessed all retrieved Pinnacle acetabular components for type, damage modes and patient demographics. The leverage force required to dissociate Pinnacle liners was also measured and compared with another commonly used acetabular system, the Trident (Stryker Orthopaedics).
The estimated minimum incidence of liner dissociation from our data was 0.35%. Characterisation of dissociated Pinnacle cases (n = 26) revealed 73% were female with an average age of 59 compared to all retrieved Pinnacle cases (n = 212) where 58% were female with an average age of 66. Retrieval analysis indicated plastic deformation of the liner into an ovoid shape, signs of impingement on the rim postero-superiorly and shearing of the liner's anti-rotation tabs was common. Mechanical testing indicated that the dissociation strength of Pinnacle cups decreases at approximately 6.6 N/year in situ (p = 0.01).
The survival rate of Pinnacle acetabular cups is exceptional with only 5% revised at 10 years. However, surgeons should be aware of the clinical symptoms and high-risk demographics when assessing patients with polyethylene Pinnacle liners.
尽管自2003年首次推出以来,品尼高髋臼髋关节系统(DePuy Synthes)已显示出出色的生存率结果,但越来越多的案例表明,品尼高衬垫可能会出现高于预期的早期分离失败率。2006年至2020年期间,我们中心从西澳大利亚医院收到了212个回收的品尼高衬垫。其中,26个因衬垫分离而被取出。
为了更好地了解这种并发症的发生频率和原因,我们评估了所有回收的品尼高髋臼组件的类型、损坏模式和患者人口统计学特征。还测量了分离品尼高衬垫所需的杠杆力,并与另一种常用的髋臼系统三叉戟(史赛克骨科)进行了比较。
根据我们的数据,衬垫分离的估计最低发生率为0.35%。对分离的品尼高病例(n = 26)的特征分析显示,73%为女性,平均年龄为59岁,而所有回收的品尼高病例(n = 212)中,58%为女性,平均年龄为66岁。取出分析表明,衬垫塑性变形为卵形、后上方边缘有撞击迹象以及衬垫防旋转凸耳剪切是常见现象。力学测试表明,品尼高髋臼杯的分离强度在原位每年大约降低6.6 N(p = 0.01)。
品尼高髋臼杯的生存率很高,10年时只有5%进行了翻修。然而,外科医生在评估使用聚乙烯品尼高衬垫的患者时,应了解其临床症状和高风险人群特征。