Malik M, Camm A J
Department of Cardiological Sciences, St. George's Hospital Medical School, London, United Kingdom.
Comput Biomed Res. 1988 Jun;21(3):289-306. doi: 10.1016/0010-4809(88)90034-1.
Like the electrocardiographic inverse problem, the object of which is the algorithmical analysis and diagnosis of the electrocardiogram (ECG), the pacemaker inverse problem is the analysis of the ECG in order to establish details of heart-pacemaker interaction (HPI) with special reference to the diagnosis of pacemaker failure. The solution to this problem is of practical importance, since it is often difficult to evaluate such records clinically. The ECG patterns of natural cardiac depolarizations and paced events may not be distinguishable and many combinations of potential pacemaker response must be taken into account. A computer system providing automatic analysis of the HPI, based on ECG data, has been developed and implemented on an IBM PC AT computer. The system uses a complex algorithm which enables the evaluation of all possible combinations of HPI events, and establishes for each of these combinations its correspondence to the specified pacemaker algorithm. The system is written in Pascal and its source text has approximately 11,000 lines. The first version of the system has been tested with algorithms of the dual chamber cardiac pulse generator AUTIMA-II (Telectronics). The interactive input of the system allows the pacemaker algorithm and the ECG, in the form of timing of "definite" and "possible" sensing events and pacemaker pulses, to be specified. The analysis establishes whether the device operates correctly within permitted error. Should one or more correct explanations of the specified case be found, the system prints simplified patterns of the ECG accompanied by a simulated marker channel tracing the possible HPIs.
如同以心电图(ECG)的算法分析与诊断为目标的心电图逆问题一样,起搏器逆问题是对心电图进行分析,以便特别针对起搏器故障的诊断来确定心脏 - 起搏器相互作用(HPI)的细节。解决这个问题具有实际重要性,因为临床上常常难以评估此类记录。自然心脏去极化和起搏事件的心电图模式可能无法区分,并且必须考虑潜在起搏器反应的许多组合。一个基于心电图数据提供HPI自动分析的计算机系统已被开发出来,并在IBM PC AT计算机上实现。该系统使用一种复杂算法,能够评估HPI事件的所有可能组合,并为这些组合中的每一个确定其与指定起搏器算法的对应关系。该系统用Pascal编写,其源文本大约有11000行。该系统的第一个版本已使用双腔心脏脉冲发生器AUTIMA - II(泰利特ronics)的算法进行了测试。该系统的交互式输入允许以“确定”和“可能”感知事件以及起搏器脉冲的时间形式指定起搏器算法和心电图。分析确定该设备在允许的误差范围内是否正确运行。如果找到指定情况的一个或多个正确解释,系统会打印出心电图的简化模式,并伴有跟踪可能的HPI的模拟标记通道。