Abstract

INTRODUCTION

The Platelet Function Analyzer-100 (PFA-100) is a point of care instrument that simulates plug formation under high shear flow. The PFA-100 measures the time required to occlude the aperture in a biochemically active cartridge and is expressed in a term of closure time (CT). In Algeria, the reference values used in clinical laboratories are of Western origin. However, ethnic, genetic, dietary environmental, and diet differences between populations may affect reference intervals. We established the reference intervals of PFA-100 closure times in healthy Algerian adults according to the International Federation of Clinical Chemistry method, and we compared them with those of Western and Asian countries.

MATERIAL AND METHODS

We enrolled 303 healthy blood donors in the study. 218 subjects met inclusion criteria. We analyzed the blood sample on the PFA-100 for CT with both the collagen epinephrine and collagen ADP cartridges.

RESULTS

The reference intervals of PFA-100 collagen epinephrine CT and PFA-100 collagen ADP CT were 91-207 seconds and 71-144 seconds, respectively. Compared to Western and Asian populations, there were significant differences. The upper limits of CTs were higher for Algerians in this study. Our findings show that many healthy Algerians would be incorrectly identified as having a primary hemostasis abnormality according to the reference intervals of the manufacturer and scientific literature.

CONCLUSION

This report provides the first reference intervals for PFA-100 CTs in healthy Algerian adults. These results improve the accuracy of diagnosis and patient care in Algeria.

摘要

简介

血小板功能分析仪-100(PFA-100)是一种即时检测仪器,可模拟高剪切流条件下的栓子形成。PFA-100 测量在生化活性管中封闭孔径所需的时间,并用闭合时间(CT)表示。在阿尔及利亚,临床实验室使用的参考值来自西方。然而,人群之间的种族、遗传、饮食环境和饮食差异可能会影响参考区间。我们根据国际临床化学联合会的方法建立了健康阿尔及利亚成年人的 PFA-100 闭合时间参考区间,并与西方和亚洲国家进行了比较。

材料和方法

我们招募了 303 名健康献血者进行研究。218 名受试者符合纳入标准。我们使用胶原肾上腺素和胶原 ADP 管在 PFA-100 上分析血液样本的 CT。

结果

PFA-100 胶原肾上腺素 CT 和 PFA-100 胶原 ADP CT 的参考区间分别为 91-207 秒和 71-144 秒。与西方和亚洲人群相比,存在显著差异。与本研究中的西方人相比,CT 值的上限更高。我们的研究结果表明,根据制造商和科学文献的参考区间,许多健康的阿尔及利亚人将被错误地识别为存在原发性止血异常。

结论

本报告提供了健康阿尔及利亚成年人 PFA-100 CT 的第一个参考区间。这些结果提高了阿尔及利亚的诊断和患者护理的准确性。