Department of Radiology and Center for Imaging Science, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea.
Medicine (Baltimore). 2021 Apr 16;100(15):e24921. doi: 10.1097/MD.0000000000024921.
This study aimed to compare the accuracy and inter- and intra-observer reproducibility of the measured elasticity between 2 shear wave elastography systems. Three breast radiologists examined 8 targets of 4 different levels of stiffness (size: 11 mm, 4 mm) in an elasticity phantom (Customized 049A Elasticity QA Phantom, CIRS, Norfolk, VA, USA) using 2 different shear wave ultrasound elastography systems: SuperSonic Imagine (SSI) (SSI, Aix en Provence, France) and ShearScan (RS-80A, Samsung Medison, Seoul, Korea). Three radiologists performed ultrasound (US) elastography examinations for the phantom lesions using 2 equipment over a 1-week interval. Intra- and inter-observer reproducibility and the accuracy of the measured elasticity were analyzed and compared between the 2 systems. The accuracy of shape was also analyzed by shape-matching between B-mode and elastography color image. Intra-class correlation coefficients (ICC) were used in statistical analysis. For measured elasticity, the intra-observer and inter-observer reproducibility were excellent in both SSI and ShearScan (0.994 and 0.998). The overall accuracy was excellent in both systems, but the accuracy in small lesions (4 mm target) was lower in SSI than ShearScan (0.780 vs 0.967). The accuracy of shape-matching on the elastography image was 59.0% and 81.4% in the SSI and ShearScan, respectively. In conclusion, the SSI and ShearScan showed excellent intra- and inter-observer reproducibility. The accuracy of the Young's modulus was high in both the SSI and ShearScan, but the SSI showed decreased accuracy in measurement of elasticity in small targets and poor shape-matching between the B-mode image and color-coded elastography image.
本研究旨在比较两种剪切波弹性成像系统测量的弹性的准确性和观察者内及观察者间可重复性。三位乳腺放射科医生使用两种不同的剪切波超声弹性成像系统(SuperSonic Imagine(SSI)和 ShearScan)检查了弹性体模(Customized 049A Elasticity QA Phantom,CIRS,Norfolk,VA,USA)中的 8 个目标(大小:11mm、4mm)的 4 个不同硬度水平。三位放射科医生在一周的时间内使用两种设备对幻影病变进行了超声(US)弹性成像检查。分析并比较了两种系统之间的测量弹性的观察者内和观察者间可重复性和准确性。还通过 B 模式和弹性成像彩色图像之间的形状匹配分析了形状的准确性。统计分析中使用了组内相关系数(ICC)。对于测量的弹性,SSI 和 ShearScan 的观察者内和观察者间可重复性均为优秀(0.994 和 0.998)。两种系统的整体准确性均为优秀,但在 SSI 中,小病变(4mm 目标)的准确性低于 ShearScan(0.780 与 0.967)。弹性成像图像上的形状匹配准确性在 SSI 和 ShearScan 中分别为 59.0%和 81.4%。总之,SSI 和 ShearScan 表现出优秀的观察者内和观察者间可重复性。两种系统的杨氏模量准确性均较高,但 SSI 在测量小目标的弹性时准确性降低,且 B 模式图像与彩色编码弹性图像之间的形状匹配较差。