Suzuki Y, Hayashi N, Sekiba K
Department of Obstetrics and Gynaecology, Okayama University Medical School, Japan.
J Chromatogr. 1988 Apr 8;426(1):33-40. doi: 10.1016/s0378-4347(00)81924-6.
An automated direct assay system using high-performance liquid chromatography was developed for the simultaneous measurement of estradiol, estrone, progesterone, 17 alpha-hydroxyprogesterone, 20 alpha-hydroxyprogesterone, testosterone and androstenedione in biological fluids. A comparison between the values measured by this method and by radioimmunoassay revealed good correlation for estradiol (r = 0.938, p less than 0.001) and progesterone (r = 0.903, p less than 0.001). Estradiol and estrone could be analysed above the level of 250 pg/ml, and progesterone, 17 alpha-hydroxyprogesterone, 20 alpha-hydroxyprogesterone, testosterone and androstenedione could be analysed above the level of 5.0-7.5 ng/ml. The method was applied to the clinical appraisal of placental function and maturation of ovarian follicles.
开发了一种使用高效液相色谱的自动化直接检测系统,用于同时测量生物体液中的雌二醇、雌酮、孕酮、17α-羟孕酮、20α-羟孕酮、睾酮和雄烯二酮。该方法与放射免疫分析法测量值之间的比较显示,雌二醇(r = 0.938,p < 0.001)和孕酮(r = 0.903,p < 0.001)具有良好的相关性。雌二醇和雌酮可在250 pg/ml以上水平进行分析,孕酮、17α-羟孕酮、20α-羟孕酮、睾酮和雄烯二酮可在5.0 - 7.5 ng/ml以上水平进行分析。该方法应用于胎盘功能和卵泡成熟的临床评估。