Indian Heart J. 2021 Mar-Apr;73(2):221-222. doi: 10.1016/j.ihj.2021.03.003. Epub 2021 Mar 17.
Radiation exposure during electrophysiology procedures has been a point of discussion. We measured the ionising radiation dosage during ablation procedures for supraventricular tachycardia. This was compared with coronary angiographies performed via the radial route to put it in perspective. We found that the radiation dosage during the ablation procedure was far lower, less than forty percent of that during coronary angiography (Air Kerma 249.1 mGy ± 266.95 mGy v/s 671.9 mGy ± 328.6 mGy; p < 0.001).
在电生理程序期间的辐射暴露一直是讨论的焦点。我们测量了用于治疗室上性心动过速的消融程序期间的电离辐射剂量,并通过经桡动脉途径进行的冠状动脉造影将其进行了比较。我们发现消融程序期间的辐射剂量要低得多,不到冠状动脉造影的百分之四十(空气比释动能 249.1 毫戈瑞±266.95 毫戈瑞 v/s 671.9 毫戈瑞±328.6 毫戈瑞;p<0.001)。