Bioanalytical Sciences and Food Analysis, University of Bayreuth, Bayreuth, Germany.
Rapid Commun Mass Spectrom. 2021 Aug 30;35(16):e9110. doi: 10.1002/rcm.9110.
High mass accuracy is indispensable for reliable identification in matrix-assisted laser desorption/ionization mass spectrometry (MALDI MS) imaging. Ubiquitous matrix ions can serve as reference masses for mass calibration if their sum formula is known. Here we report an overview of ions generated on tissue by 11 common MALDI matrices for use in internal or external mass calibration.
Matrices covered in this study were applied onto coronal mouse brain sections using a pneumatic sprayer setup. MALDI imaging was performed on a Q Exactive HF orbital trapping mass spectrometer coupled to an AP-SMALDI 10 source. Measurements were conducted with high mass resolution (240 k full width at half maximum at m/z 200) and high mass accuracy with a root mean square mass error of better than 1.5 ppm achieved via internal mass calibration using matrix ions.
MALDI MS imaging was used to investigate ions generated on tissue by 11 common MALDI matrices. An example of using matrix ions for internal mass calibration in MALDI imaging of drug substances and lipids in murine lung sections is presented. Tables containing the cluster composition, sum formulae, and the measured and theoretical m/z ratios of the identified ions were compiled for each matrix.
Using matrix ions as reference masses for internal and external mass calibration in MALDI MS imaging is an effective and elegant way to achieve sub-ppm mass accuracy as it makes use of ubiquitous signals present in every MALDI MS spectrum without the need for an additional calibration standard.
在基质辅助激光解吸电离质谱(MALDI MS)成像中,高质量精度对于可靠的识别是必不可少的。如果已知基质离子的分子式,它们可以作为参考质量用于质量校准。在这里,我们报告了在组织上由 11 种常见 MALDI 基质产生的离子的概述,用于内部或外部质量校准。
本研究中涵盖的基质使用气动喷雾器装置施加到冠状小鼠脑切片上。MALDI 成像在与 AP-SMALDI 10 源耦合的 Q Exactive HF 轨道阱质谱仪上进行。测量是在高质量分辨率(240 k 在 m/z 200 处的半峰全宽)和高质量精度下进行的,通过使用基质离子进行内部质量校准,实现了优于 1.5 ppm 的均方根质量误差。
使用 MALDI MS 成像来研究由 11 种常见 MALDI 基质在组织上产生的离子。展示了在使用基质离子进行内部质量校准的情况下,对鼠肺切片中的药物物质和脂质进行 MALDI 成像的示例。为每个基质编制了包含簇组成、分子式以及鉴定离子的实测和理论 m/z 比的表格。
使用基质离子作为 MALDI MS 成像中内部和外部质量校准的参考质量,是一种有效且优雅的方法,可以实现亚 ppm 的质量精度,因为它利用了每个 MALDI MS 光谱中存在的无处不在的信号,而无需额外的校准标准。