Technical Note: Volumetric coverage in breast tomosynthesis images - Phantom QC results from the TMIST study.

作者信息

Maki Aili K, Mainprize James G, Pisano Etta D, Mawdsley Gordon E, Skarpathiotakis Mia, Yaffe Martin J

机构信息

Physical Sciences, Sunnybrook Research Institute, 2075 Bayview Avenue, Toronto, ON, M4N 3M5, Canada.

Department of Radiology, Beth Israel Deaconess Medical Center, Boston, MA, USA.

出版信息

Med Phys. 2021 Jul;48(7):3623-3629. doi: 10.1002/mp.14911. Epub 2021 May 17.

Abstract

PURPOSE

In the reconstruction of volume breast images from x-ray projections in breast tomosynthesis, some tomographic systems truncate the image data presented to the radiologist such that a non-negligible amount of tissue may be missing from the breast image. QC tests were conducted to determine if this problem existed in imaging in the TMIST study.

METHODS

Test tools developed for TMIST containing small objects at known heights were used in routine weekly and annual QC testing of tomosynthesis units to assess the degree to which phantom material that was irradiated in imaging was excluded from the reconstructed image. Results from 318 tests on five system types from three manufacturers are reported.

RESULTS

The presence and extent of this problem varied among system types. The cause was most frequently related to machine errors in the determination of breast thickness or to deflection of components during breast compression. In particular, the problem occurred when a compression paddle other than the one calibrated for tomosynthesis was used for the tests. This was also verified to have occurred in some clinical imaging.

CONCLUSIONS

Missing volume can be avoided by intentionally reconstructing additional image slices above and below the presumed locations of the breast support and compression plate. A compression paddle which has been calibrated for tomosynthesis should be used both for clinical imaging and testing. The prevalence of this phenomenon suggests that more frequent testing for volume coverage may be advisable.

摘要

目的

在乳腺断层合成中从X射线投影重建乳腺体积图像时,一些断层扫描系统会截断呈现给放射科医生的图像数据,以至于乳腺图像中可能会遗漏数量不可忽视的组织。进行了质量控制测试,以确定TMIST研究中的成像是否存在此问题。

方法

为TMIST开发的包含已知高度小物体的测试工具用于断层合成设备的常规每周和年度质量控制测试,以评估成像中被照射的体模材料被排除在重建图像之外的程度。报告了对来自三个制造商的五种系统类型进行的318次测试的结果。

结果

该问题的存在和程度因系统类型而异。原因最常与确定乳腺厚度时的机器误差或乳腺压缩过程中部件的偏斜有关。特别是,当使用除为断层合成校准的压缩板之外的其他压缩板进行测试时,就会出现这个问题。这也被证实在一些临床成像中发生过。

结论

通过有意在乳腺支撑和压缩板的假定位置上方和下方重建额外的图像切片,可以避免体积遗漏。应为临床成像和测试使用已针对断层合成校准的压缩板。这种现象的普遍性表明,可能建议更频繁地进行体积覆盖测试。