Abstract

BACKGROUND

Total knee arthroplasty is a very successful standard treatment for severe osteoarthritis. Nevertheless, the literature reports tibial debonding between implant and bone cement as well as radiolucent lines related to the tibial components of different knee systems. Regardless of cementing techniques and the influences during surgery, we examined the design of a newly developed knee system and its predecessors (Attune, Attune S+, P.F.C. Sigma, P.F.C. Sigma RP/M.B.T., all DePuy).

METHODS

We investigated the dimensions of the tibial components and the fit between them and their bone bed after instrumentation in a foam material.

RESULTS

Our results showed considerable differences for the used knee prostheses as well as their tibial instrumentation options with a corresponding risk for incomplete seating.

CONCLUSION

The orthopedic surgeons need to be aware of these design features and the resulting increased seating resistance especially in hard and sclerotic bone.

ARTICLE FOCUS

Comparison of the tibial instruments and the different design options of the Attune knee system and its predecessor knee prostheses.

KEY MESSAGES

The Attune implant showed incomplete seating because of too much press fit and an uneven bone quality or sclerosis can result in tilting of the tibial component.

STRENGTHS AND LIMITATIONS

This is the first study investigating the Attune knee and its predecessor in terms of implant seating and press fit. The foam material is a limitation.

摘要

背景

全膝关节置换术是治疗严重骨关节炎的非常成功的标准治疗方法。然而,文献报道了不同膝关节系统的胫骨组件与骨水泥之间存在的植入物和骨水泥之间的胫骨脱粘以及与胫骨组件相关的透亮线。无论使用何种粘结技术以及手术过程中的影响如何,我们都检查了新开发的膝关节系统及其前代产品(Attune、Attune S+、PFC Sigma、PFC Sigma RP/MBT,均为 DePuy)的设计。

方法

我们在泡沫材料中研究了胫骨组件的尺寸以及它们之间的配合及其骨床在仪器化后的情况。

结果

我们的研究结果显示,所使用的膝关节假体及其胫骨仪器化选项之间存在明显差异,存在不完全就位的风险。

结论

骨科医生需要意识到这些设计特点以及由此导致的增加的就位阻力,特别是在坚硬和硬化的骨中。

文章重点

比较 Attune 膝关节系统及其前代膝关节假体的胫骨器械和不同设计选项。

关键信息

Attune 植入物由于过盈配合和不均匀的骨质量或硬化而导致不完全就位,可能导致胫骨组件倾斜。

优势和局限性

这是第一项研究,研究了 Attune 膝关节及其前代产品在植入物就位和压配合方面的情况。泡沫材料是一个限制因素。