Karlik S J, Heatherley T, Pavan F, Stein J, Lebron F, Rutt B, Carey L, Wexler R, Gelb A
Department of Diagnostic Radiology, University of Western Ontario, London, Canada.
Magn Reson Med. 1988 Jun;7(2):210-21. doi: 10.1002/mrm.1910070209.
This paper describes the facilities successfully used to provide patient monitoring and anesthesia support in a 1.5-T imaging installation. The requirements for the MRI site for anesthetic gases, shielded power, and radiofrequency ports are outlined. Specific modifications in anesthesia machines, anesthesia cart, laryngoscope, mercury sphygmomanometer, oximeter, and remote blood pressure devices are described. Additional aspects of patient monitoring and support, e.g., electrocardiogram and pressure infusion, are also discussed.
本文描述了在1.5-T成像设备中成功用于提供患者监测和麻醉支持的设施。概述了MRI场地对麻醉气体、屏蔽电源和射频端口的要求。描述了麻醉机、麻醉推车、喉镜、水银血压计、血氧仪和远程血压设备的具体改进。还讨论了患者监测和支持的其他方面,如心电图和压力输注。