Validation of the QMon-20 oscillometric blood pressure monitor for professional office use in the general population according to the ANSI/ESH/ISO 81060-2:2018 protocol.

机构信息

Nursing Department, Nursing and Occupational Therapy College, University of Extremadura, Cáceres.

Medicine Department, Hospital Clínico de Madrid, Madrid.

出版信息

Blood Press Monit. 2021 Oct 1;26(5):393-395. doi: 10.1097/MBP.0000000000000540.

Abstract

OBJECTIVE

we assessed the accuracy of the QMon-20 oscillometric upper-arm cuff device professional for office blood pressure (BP) in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018).

METHODS

Subjects were recruited according to AAMI/ESH/ISO Universal Standard in general population using the same arm sequential BP measurement method. Two cuffs of the test device were used for arm circumference 22-31 (medium) and 32-42 cm (large).

RESULTS

One-hundred and fourteen subjects were recruited and 106 were analyzed. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 0.8 ± 5.4/-0.5 ± 4.2 mmHg (systolic/diastolic). For criterion 2, the SD of the mean BP differences between the test device and reference BP per subject was 4.34/3.48 mmHg (systolic/diastolic).

CONCLUSION

The QMon-20 oscillometric device for office BPs measurement fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in general population and can be recommended for clinical use.

摘要

目的

根据医疗器械促进协会/欧洲高血压学会/国际标准化组织(AAMI/ESH/ISO)通用标准(ISO 81060-2:2018),评估 QMon-20 腕式血压袖带在一般人群中用于诊室血压(BP)测量的准确性。

方法

根据 AAMI/ESH/ISO 通用标准,使用相同的手臂顺序血压测量方法,在一般人群中招募受试者。该测试设备使用两种尺寸的袖带,臂围为 22-31cm(中号)和 32-42cm(大号)。

结果

共招募了 114 名受试者,其中 106 名进行了分析。对于验证标准 1,测试设备与参考血压读数之间的差异平均值±标准差为 0.8±5.4/-0.5±4.2mmHg(收缩压/舒张压)。对于标准 2,每位受试者的测试设备与参考 BP 之间的平均 BP 差异的标准差为 4.34/3.48mmHg(收缩压/舒张压)。

结论

QMon-20 腕式血压计用于诊室血压测量符合 AAMI/ESH/ISO 通用标准(ISO 81060-2:2018)的所有要求,可推荐用于临床使用。