Blood Press Monit. 2024 Apr 1;29(2):99-102. doi: 10.1097/MBP.0000000000000684. Epub 2023 Nov 15.
To evaluate the blood pressure (BP) measurement accuracy of the Braun BUA4000 automated oscillometric upper-arm cuff device for self-home use in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) and its Amendment 1.2020-01.
Participants were recruited to fulfill the age, sex, BP and limb circumference distribution criteria of the AAMI/ESH/ISO Universal Standard in a general population using the same arm sequential BP measurement method. A single wide-range cuff of the test device was used for arm circumference range 22-42 cm.
A total of 98 individuals were recruited and 85 were analyzed [mean age 60.3 ± 16.1 (SD) years, 44 men, arm circumference 31.5 ± 5.1 cm, range 22-41.5 cm]. For validation criterion 1, the mean difference ± SD between the test device and reference BP readings ( N = 255) was 0.9 ± 6.4/-0.3 ± 6.4 mmHg (systolic/diastolic; threshold ≤5 ± 8 mmHg). For criterion 2, the SD of the averaged BP differences between the test device and reference BP per individual ( N = 85) was 5.15/5.81 mmHg (systolic/diastolic; threshold ≤6.88/6.95 mmHg).
The Braun BUA4000 automated oscillometric BP monitor fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard in a general population and can be recommended for self-monitoring of BP by patients at home.
根据医疗器械促进协会/欧洲高血压学会/国际标准化组织(AAMI/ESH/ISO)通用标准(ISO 81060-2:2018)及其 2020-01 年修正案 1,评估 Braun BUA4000 自动示波法上臂袖带式电子血压计在一般人群中自我家庭使用的血压(BP)测量准确性。
采用相同的手臂顺序 BP 测量方法,根据 AAMI/ESH/ISO 通用标准的年龄、性别、BP 和四肢周长分布标准,招募参与者以满足一般人群的要求。测试设备的单个宽范围袖带用于臂围范围 22-42cm。
共招募了 98 人,其中 85 人进行了分析[平均年龄 60.3±16.1(SD)岁,44 名男性,臂围 31.5±5.1cm,范围 22-41.5cm]。对于验证标准 1,测试设备与参考 BP 读数( N=255)之间的平均差值±SD 为 0.9±6.4/-0.3±6.4mmHg(收缩压/舒张压;阈值≤5±8mmHg)。对于标准 2,每个个体的测试设备和参考 BP 之间平均 BP 差值的 SD( N=85)为 5.15/5.81mmHg(收缩压/舒张压;阈值≤6.88/6.95mmHg)。
Braun BUA4000 自动示波法 BP 监测器完全满足一般人群中 AAMI/ESH/ISO 通用标准的所有要求,可推荐患者在家中进行自我 BP 监测。