Blood Press Monit. 2022 Aug 1;27(4):276-279. doi: 10.1097/MBP.0000000000000600. Epub 2022 Apr 18.
To evaluate the accuracy of the EDAN SA-10 oscillometric upper-arm professional office ambulatory blood pressure (BP) monitor in general population according to the Association for the Advancement of Medical Instrumentation (AAMI)/European Society of Hypertension (ESH)/International Organization for Standardization (ISO) Universal Standard (ISO 81060-2:2018).
Subjects were recruited according to the AAMI/ESH/ISO Universal Standard using the same arm sequential BP measurement method. Four cuffs of the test device were used for arm circumference 16-21.5 cm (extra small), 20.5-28 cm (small), 27-35 cm (medium), and 34-43 cm (large).
A total of 105 subjects were recruited, and 97 subjects were included in the final analysis. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was -0.59 ± 4.04/-1.79 ± 4.39 mmHg (systolic/diastolic). For criterion 2, the SD of the averaged BP differences between the test device and reference BP per subject was 3.10/3.80 mmHg (systolic/diastolic).
The EDAN SA-10 upper-arm ambulatory BP monitor has passed all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in general population and can be recommended for clinical use.
根据医疗器械促进协会(AAMI)/欧洲高血压学会(ESH)/国际标准化组织(ISO)通用标准(ISO 81060-2:2018)评估 EDAN SA-10 臂式电子血压计在一般人群中的准确性。
采用 AAMI/ESH/ISO 通用标准,按照相同的手臂顺序血压测量方法招募受试者。该测试设备使用了四个袖带,适用于臂围 16-21.5cm(特小号)、20.5-28cm(小号)、27-35cm(中号)和 34-43cm(大号)。
共招募了 105 名受试者,最终有 97 名受试者纳入了最终分析。对于验证标准 1,测试设备和参考血压读数之间的差异的平均值±SD 为-0.59±4.04/-1.79±4.39mmHg(收缩压/舒张压)。对于标准 2,每个受试者的测试设备和参考血压之间的平均 BP 差异的 SD 为 3.10/3.80mmHg(收缩压/舒张压)。
EDAN SA-10 臂式动态血压监测仪通过了 AAMI/ESH/ISO 通用标准(ISO 81060-2:2018)的所有要求,可推荐用于临床使用。