Department of Chemistry, Sapienza University, Piazzale Aldo Moro 5, 00185 Rome, Italy.
Institute of Atmospheric Pollution Research, National Research Council (CNR), Via Salaria Km 29,300, Monterotondo St., 00015 Rome, Italy.
Food Chem. 2021 Oct 30;360:130027. doi: 10.1016/j.foodchem.2021.130027. Epub 2021 May 9.
The accurate determination of trace elements in vegetable edible oils is still an analytical challenge, owing to their low concentration levels and the complex matrix of the vegetable oils. The aim of this study was to develop a fast and simple analytical method to quantify 45 elements in small mass samples (0.5 g) of extra virgin olive oils by inductively coupled plasma mass spectrometry. To evaluate the best and fastest sample preparation procedure, ultrasonic extraction and wet digestion methods were compared using oil certified reference material with different reagent mixtures, reagent volumes, and times for sample extraction or digestion. The use of 5 mL reagent mixture F [10% (v/v) HNO and HO, 2:1 (v/v)] for sample digestion in a water bath (95 °C, 40 min) was found to produce satisfactory results in all cases as validated from sample recovery experiments over three different extra virgin olive oil samples.
准确测定植物油中的微量元素仍然是一个分析挑战,因为它们的浓度水平低,且植物油的基质复杂。本研究旨在开发一种快速、简单的分析方法,通过电感耦合等离子体质谱法对小质量样品(0.5g)的特级初榨橄榄油中的 45 种元素进行定量分析。为了评估最佳和最快的样品制备程序,比较了超声提取和湿法消解方法,使用不同试剂混合物、试剂体积和样品提取或消解时间的油认证参考物质。发现使用 5mL 试剂混合物 F(10%(v/v)HNO 和 H O,2:1(v/v))在水浴(95°C,40min)中进行样品消解,在所有情况下都能得到令人满意的结果,这通过对三个不同特级初榨橄榄油样品的样品回收实验进行了验证。