Abstract

Following a moving hot spot in the projections of raw images and profound perfusion defects in myocardial perfusion single-photon emission computed tomography (SPECT) imaging of a patient, a hypothesis was postulated that the perfusion defects were artifactual, and the high activity concentration of the gallbladder may be a culprit for this phenomenon, owing to flawed event positioning function of the gamma camera due to a malfunctioning digital event processor electronics board. To depict the characteristics of this artifact, a point source containing an activity of 3 mCi of pertechnetate is placed on the scanning table with the detector facing the table (at a distance of 30 cm), and then, in other detector positions and 1-min static images are acquired accordingly. The ratio is calculated as follows: count of the artifactual focus: 1860, count of the index focus: 705,727, and artifactual-to-index focus ratio: 0.003. In testing the uniformity of gamma camera based on the National Electrical Manufacturers Association protocol, a nonuniform response was detected, seemingly, a smaller field of view (FOV) is reproduced in the main FOV causing nonuniformity more than the acceptable level. The smaller flood image lies in the upper right corner of the main flood image. In essence, the extremely bright gallbladder was the source of error, and its image was reproduced in the FOV, which was superimposed on the left ventricular myocardium in some of the projections and was propagated to SPECT images.

摘要
https://cdn.ncbi.nlm.nih.gov/pmc/blobs/94f0/8130697/d53ae0a315ea/IJNM-36-69-g001.jpg

相似文献

3
Intrinsic uniformity requirements for pinhole SPECT.
J Nucl Med Technol. 2006 Mar;34(1):43-7.

引用本文的文献

1
"Picture-in-Picture" Artifact in Post-therapeutic I Whole-body Survey: Deceiving Spot View but Unraveling Whole-body Scanning.
Mol Imaging Radionucl Ther. 2022 Jun 27;31(2):142-144. doi: 10.4274/mirt.galenos.2021.65882.

本文引用的文献

2
Contamination, a Major Problem in Nuclear Medicine Imaging: How to Investigate, Handle, and Avoid It.
J Nucl Med Technol. 2017 Sep;45(3):241-242. doi: 10.2967/jnmt.117.190256. Epub 2017 Apr 13.
3
Routine quality control recommendations for nuclear medicine instrumentation.
Eur J Nucl Med Mol Imaging. 2010 Mar;37(3):662-71. doi: 10.1007/s00259-009-1347-y.
4
Routine quality control of clinical nuclear medicine instrumentation: a brief review.
J Nucl Med. 2008 Jul;49(7):1114-31. doi: 10.2967/jnumed.107.050203.
5
The day of contamination.
J Nucl Med Technol. 2003 Mar;31(1):21-4.
6
Instrument- and computer-related problems and artifacts in nuclear medicine.
Semin Nucl Med. 1996 Oct;26(4):256-77. doi: 10.1016/s0001-2998(96)80003-1.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验