使用拟人化躯干模型和水袋模拟软组织和乳腺衰减,对新型碲化镉锌相机与传统单光子发射计算机断层扫描/计算机断层扫描(SPECT/CT)进行性能比较。

A performance comparison of novel cadmium-zinc-telluride camera and conventional SPECT/CT using anthropomorphic torso phantom and water bags to simulate soft tissue and breast attenuation.

作者信息

Liu Chia-Ju, Cheng Ju-Shin, Chen Yi-Chieh, Huang Yih-Hwen, Yen Ruoh-Fang

机构信息

Department of Nuclear Medicine, National Taiwan University Hospital, No. 7, Chung-Shan South Road, Taipei, 100, Taiwan.

出版信息

Ann Nucl Med. 2015 May;29(4):342-50. doi: 10.1007/s12149-015-0952-z. Epub 2015 Jan 28.

Abstract

PURPOSES

This study was aimed to compare the physical performances of cadmium-zinc-telluride (CZT) camera and conventional Anger camera. An anthropomorphic torso phantom and water bags to simulate breasts were used to evaluate artifacts arising from soft tissue attenuation.

MATERIALS AND METHODS

Linear source studies were performed to evaluate extrinsic resolution of CZT camera (Discovery NM 530c, GE) and conventional single-photon emission computed tomography (SPECT) Anger camera (Symbia T2, Siemens). Three sets of phantom experiments: cardiac phantom only (phantom H), anthropomorphic torso phantom added (phantom T), and torso phantom with water bags attached (phantom B), with Tc-99m were performed on both scanners. Imaging performances were evaluated through count sensitivity, contrast-to-noise ratio, quantitative sharpness profile, wall thickness, perfusion uniformity (measured by standard deviation of perfusion percentage of 20 segments using quantitative perfusion SPECT (QPS) software, Cedars-Sinai), and visual imaging quality (using 20-segment sum defect scores (SDS) of QPS) for CZT camera, conventional SPECT without and with computed tomography transmission attenuation correction (AC).

RESULTS

CZT cameras had higher extrinsic resolution than conventional SPECT. Myocardium count sensitivity of CZT camera is about threefold of conventional SPECT. Contrast-to-noise ratios and sharpness profiles are higher on CZT camera but degraded while extracardiac soft tissue presented. Myocardial walls measured on CZT images were thicker. Images of CZT had lower SDS, while AC reduced the differences of SDS between CZT and CC. Perfusion images from CZT had the better uniformity than SPECT without or with AC. Breast attenuation was less prominent on CZT camera than conventional SPECT, while inferior and inferolateral segments still suffer marked soft tissue attenuation on CZT camera.

CONCLUSIONS

CZT camera has better physical performance and image quality with less artificial perfusion defects than conventional SPECT. CZT camera also has less breast attenuation than conventional SPECT. However, extracardiac soft tissue may degrade the superior performance of CZT camera, and attenuation correction methods are still needed to solve the attenuation issues in inferior and inferolateral myocardium.

摘要

目的

本研究旨在比较碲锌镉(CZT)相机与传统安格相机的物理性能。使用拟人化躯干模型和模拟乳房的水袋来评估软组织衰减引起的伪影。

材料与方法

进行线性源研究以评估CZT相机(GE公司的Discovery NM 530c)和传统单光子发射计算机断层扫描(SPECT)安格相机(西门子公司的Symbia T2)的外在分辨率。在两台扫描仪上对三组模型实验进行了研究:仅心脏模型(模型H)、添加拟人化躯干模型(模型T)以及附有模拟乳房水袋的躯干模型(模型B),均使用锝-99m。通过计数灵敏度、对比噪声比、定量锐度剖面图、心肌壁厚度、灌注均匀性(使用Cedars-Sinai的定量灌注SPECT(QPS)软件测量20个节段灌注百分比的标准差)以及视觉成像质量(使用QPS的20节段总和缺陷评分(SDS))对CZT相机、未进行和进行了计算机断层扫描透射衰减校正(AC)的传统SPECT的成像性能进行评估。

结果

CZT相机的外在分辨率高于传统SPECT。CZT相机的心肌计数灵敏度约为传统SPECT的三倍。CZT相机上的对比噪声比和锐度剖面图更高,但在存在心外软组织时会降低。在CZT图像上测量的心肌壁更厚。CZT相机的图像具有更低的SDS,而AC减少了CZT相机与传统相机之间SDS的差异。CZT相机的灌注图像比未进行或进行了AC的SPECT具有更好的均匀性。与传统SPECT相比,CZT相机上的乳房衰减不太明显,但在CZT相机上心尖和下侧壁节段仍受到明显的软组织衰减。

结论

与传统SPECT相比,CZT相机具有更好的物理性能和图像质量,且人工灌注缺陷更少。CZT相机的乳房衰减也比传统SPECT少。然而,心外软组织可能会降低CZT相机的优越性能,并且仍需要衰减校正方法来解决心尖和下侧壁心肌的衰减问题。