Matsumoto N, Berman D S, Kavanagh P B, Gerlach J, Hayes S W, Lewin H C, Friedman J D, Germano G
Division of Nuclear Medicine, Department of Imaging, Cedars-Sinai Mediical Center, University of California Los Angeles School of Medicine, 90048, USA.
J Nucl Med. 2001 May;42(5):687-94.
Patient motion during myocardial perfusion SPECT can produce images that show artifactual perfusion defects. The relationship between the degree of motion and the extent of artifactual perfusion defects is not clear for either single- or double-head detectors. Using both single- and double-head detectors and quantitative perfusion SPECT (QPS) software, we studied the pattern and extent of defects induced by simulated motion and validated a new automatic motion-correction program for myocardial perfusion SPECT.
Vertical motion was simulated by upward shifting of the raw projection datasets in a returning pattern (bounce) and in a nonreturning pattern at 3 different phases of the SPECT acquisition (early, middle, and late), whereas upward creep was simulated by uniform shifting throughout the acquisition. Lateral motion was similarly simulated by left shifting of the raw projection datasets in a returning pattern and in a nonreturning pattern. Simulations were performed using single- and double-head detectors, and simulated motion was applied to projection images from 8 patients who had normal 99mTc-sestamibi SPECT findings. Additionally, images from 130 patients with actual clinical motion were assessed before and after motion correction. The extent of perfusion defects was assessed by QPS, and a 20-segment, 5-point scoring system was used to assess the effect of motion on the presence and extent of perfusion defects.
Of 12 bounce simulations, the bouncing motion failed to produce significant (>3%) perfusion defects with either the single- or the double-head detector. With the single-head detector, early shifting created the largest defect, whereas with the double-head detector, shifting during the middle of the acquisition created the largest defect. With regard to upward creep, defects were of larger extent with the double- than the single-head detector. With the single-head detector, 8 of 20 simulated motion patterns yielded significant perfusion defects of the left ventricle, 7 (88%) of which were significantly improved after motion correction. With the double-head detector, 12 of 20 patterns yielded significant defects, all of which improved significantly after correction. Of 2,600 segments in the 130 patients with actual clinical motion, only 1.3% (30/2,259) of segments that were considered normal (score = 0 or 1) changed to abnormal (score = 2-4) after motion correction, whereas 27% (92/341) of abnormal segments were reclassified as normal after motion correction.
Artifactual perfusion defects created by simulated motion are a function of the time, degree, and type of motion and the number of camera detectors. Application of an automatic motion-correction algorithm effectively decreases motion artifacts on myocardial perfusion SPECT images.
心肌灌注单光子发射计算机断层扫描(SPECT)过程中患者的运动可产生显示伪灌注缺损的图像。对于单探头或双探头探测器,运动程度与伪灌注缺损范围之间的关系尚不清楚。我们使用单探头和双探头探测器以及定量灌注SPECT(QPS)软件,研究了模拟运动引起的缺损模式和范围,并验证了一种用于心肌灌注SPECT的新型自动运动校正程序。
在SPECT采集的3个不同阶段(早期、中期和晚期),通过以返回模式(弹跳)和非返回模式向上移动原始投影数据集来模拟垂直运动,而通过在整个采集过程中均匀移动来模拟向上蠕动。横向运动同样通过以返回模式和非返回模式向左移动原始投影数据集来模拟。使用单探头和双探头探测器进行模拟,并将模拟运动应用于99mTc-司他米比SPECT检查结果正常的8例患者的投影图像。此外,对130例实际存在临床运动的患者的图像在运动校正前后进行评估。通过QPS评估灌注缺损的范围,并使用20节段、5分评分系统评估运动对灌注缺损的存在和范围的影响。
在12次弹跳模拟中,无论是单探头还是双探头探测器,弹跳运动均未产生显著(>3%)的灌注缺损。对于单探头探测器,早期移动产生的缺损最大,而对于双探头探测器,采集中期的移动产生的缺损最大。关于向上蠕动,双探头探测器产生的缺损范围比单探头探测器更大。对于单探头探测器,20种模拟运动模式中有8种产生了左心室的显著灌注缺损,其中7种(88%)在运动校正后有显著改善。对于双探头探测器,20种模式中有12种产生了显著缺损,所有这些在校正后均有显著改善。在130例实际存在临床运动的患者的2600个节段中,运动校正后,只有1.3%(30/2259)被认为正常(评分=0或1)的节段变为异常(评分=2 - 4),而27%(92/341)的异常节段在运动校正后被重新分类为正常。
模拟运动产生的伪灌注缺损是运动时间、程度、类型以及相机探测器数量的函数。应用自动运动校正算法可有效减少心肌灌注SPECT图像上的运动伪影。