Suppr超能文献

当植入式心脏除颤器可能危及患者生命时!

When an Implantable Cardioverter Defibrillator can Kill your Patient!

机构信息

Instituto de Ritmologia Cardíaca, Santa Catarina, Brazil.

Serviço de Eletrofisiologia e Arritmia Cardíaca, Centro Hospitalar UNIMED, Santa Catarina, Brazil.

出版信息

Braz J Cardiovasc Surg. 2021 Apr 1;36(2):268-271. doi: 10.21470/1678-9741-2019-0219.

Abstract

Inappropriate therapy due to noise oversensing caused a true ventricular fibrillation (VF) and a life-threatening event in a patient. A 19-year-old patient with surgically corrected congenital heart disease and systolic dysfunction had an implantable cardioverter defibrillator implanted for primary prevention in 2013. This patient was admitted at the Emergency Department in June 2018 after receiving eight shocks from the device on the same day, with a prolonged syncope after the third shock. Another noise-induced VF detection occurred, and two inappropriate shocks followed sequentially, causing true VF. Four appropriate shocks were subsequently needed until sinus rhythm was finally restored.

摘要

由于噪声过度感知导致的不恰当治疗,引发了一位患者的真正心室颤动(VF)和危及生命的事件。一位 19 岁的患者患有先天性心脏病和收缩功能障碍,并于 2013 年接受了植入式心脏复律除颤器植入以进行一级预防。该患者于 2018 年 6 月因同日接受了该设备的 8 次电击而被收入急诊室,在第三次电击后长时间出现晕厥。另一次因噪声引起的 VF 检测发生,随后连续发生了两次不恰当的电击,导致真正的 VF。随后需要进行四次适当的电击,直到最终恢复窦性节律。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/0c6d/8163278/a859d9fc46fc/rbccv-36-02-0268-g01.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验