The displacements of stapes piston prostheses, induced by variations of static air pressure from 0 to +/- 400 mm H2O, were examined in 9 temporal bone preparations by a microscopic measuring method. In the normal middle ear, the gliding incudo-malleal joint, serving as a protective mechanism, confined these movements to a mean of 232 microns. These in- and excursions could exceed 0.5 mm, however, with an ankylosed incudo-malleal joint. Arthrosis of this joint is not rare in otosclerosis. In view of these excessive displacements, compared to inner ear dimensions, piston prostheses should always dip into the vestibulum in the lower half of the footplate, far from endolymphatic membranes. A simple technique may intraoperatively reveal the gliding or ankylosed condition of incudo-malleal joint surfaces, prior to adapting the length of the prosthesis.
采用显微测量法,在9个颞骨标本中检测了镫骨活塞型人工听骨在0至±400 mm H₂O静态气压变化时的位移情况。在正常中耳中,作为一种保护机制,滑动的砧镫关节将这些运动限制在平均232微米。然而,当砧镫关节强直时,这些进出运动可能超过0.5毫米。这种关节的关节炎在耳硬化症中并不罕见。鉴于与内耳尺寸相比这些过度位移,活塞型人工听骨应始终置于足板下半部分的前庭内,远离内淋巴膜。在调整人工听骨长度之前,一种简单的技术可以在术中揭示砧镫关节面的滑动或强直状态。