Lyophilized calf uterine tissue cytosol standards for estradiol receptor (ER) and progesterone receptor (PR), and freeze-dried tissue powders for PR determination were prepared. For the ER assay 4 standards were produced with no, low, medium, and high ER levels. The ER binding capacities (mean +/- SD) were low--261 +/- 46 (Brno) and 221 +/- 64 (Budapest), medium--451 +/- 100 and 340 +/- 59, high--712 +/- 139 and 581 +/- 102 fmol/mg protein, respectively. Two receptor-positive tissue powders and two positive cytosol standards in lyophilized form were used for the PR assay with receptor contents of 110, 148, 787, and 786 fmol/mg assayed in Brno, and 148, 250, 693, and 671 fmol/mg protein determined in Budapest. The stability of the standards was good for 10 months. All parameters tested in the two laboratories (specific binding capacities, Kd values, variation coefficients of the results) were in good correlation. Lyophilized calf uterine cytosol standards for ER and PR, and freeze-dried tissue powders for PR proved to be suitable materials for interlaboratory quality control of steroid receptor determination.
制备了用于雌二醇受体(ER)和孕酮受体(PR)检测的冻干小牛子宫组织胞浆标准品,以及用于PR测定的冻干组织粉末。对于ER检测,制备了4种标准品,其ER水平分别为无、低、中、高。ER结合能力(平均值±标准差)分别为低水平——布尔诺为261±46、布达佩斯为221±64;中等水平——451±100和340±59;高水平——712±139和581±102 fmol/mg蛋白质。两种受体阳性组织粉末和两种冻干形式的阳性胞浆标准品用于PR检测,布尔诺检测的受体含量分别为110、148、787和786 fmol/mg,布达佩斯测定的为148、250、693和671 fmol/mg蛋白质。标准品在10个月内稳定性良好。两个实验室检测的所有参数(特异性结合能力、Kd值、结果变异系数)均具有良好的相关性。用于ER和PR检测的冻干小牛子宫胞浆标准品以及用于PR检测的冻干组织粉末被证明是类固醇受体测定实验室间质量控制的合适材料。