两种屏-片组合在乳腺接触式摄影和放大摄影中的比较:微钙化的可检测性

Comparison of two screen-film combinations in contact and magnification mammography: detectability of microcalcifications.

作者信息

Oestmann J W, Kopans D B, Linetsky L, Hall D A, McCarthy K A, White G, Swann C, Kelley J E, Johnson L L

机构信息

Department of Radiology, Massachusetts General Hospital, Boston.

出版信息

Radiology. 1988 Sep;168(3):657-9. doi: 10.1148/radiology.168.3.3406394.

Abstract

A new dual-screen, dual-emulsion-film combination that allows a decrease in radiation dose of approximately 66% was compared with a widely used single-screen, single-emulsion-film system in contact and magnification mammography. Clustered microcalcifications randomly superimposed on a breast phantom were detected, and the location and number of individual calcifications were determined by four observers. The detectability of calcifications, determined with a receiver operating characteristic (ROC) analysis area, was 0.92 for magnification and 0.82 for contact mammography with the single-emulsion-film system, compared with 0.84 and 0.72, respectively, with the dual-emulsion-film system. More clusters were correctly located and more individual calcifications were counted with magnification than with contact mammography. The dual-emulsion-film system with the magnification technique performs as well as the single-emulsion-film system with the contact technique, while retaining a decrease in required dose of approximately 40%.

摘要

在接触式和放大式乳腺摄影中,将一种新型双屏双乳剂胶片组合(可使辐射剂量降低约66%)与广泛使用的单屏单乳剂胶片系统进行了比较。对随机叠加在乳腺模型上的簇状微钙化进行检测,并由四名观察者确定单个钙化的位置和数量。通过接受者操作特征(ROC)分析区域确定钙化的可检测性,单乳剂胶片系统在放大式乳腺摄影中的可检测性为0.92,在接触式乳腺摄影中为0.82,而双乳剂胶片系统在放大式乳腺摄影和接触式乳腺摄影中的可检测性分别为0.84和0.72。与接触式乳腺摄影相比,放大式乳腺摄影能更准确地定位更多的簇状钙化,并能计数更多的单个钙化。双乳剂胶片系统结合放大技术的表现与单乳剂胶片系统结合接触技术的表现相当,同时所需剂量降低了约40%。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验