Oestmann J W, Kopans D B, Linetsky L, Hall D A, McCarthy K A, White G, Swann C, Kelley J E, Johnson L L
Department of Radiology, Massachusetts General Hospital, Boston.
Radiology. 1988 Sep;168(3):657-9. doi: 10.1148/radiology.168.3.3406394.
A new dual-screen, dual-emulsion-film combination that allows a decrease in radiation dose of approximately 66% was compared with a widely used single-screen, single-emulsion-film system in contact and magnification mammography. Clustered microcalcifications randomly superimposed on a breast phantom were detected, and the location and number of individual calcifications were determined by four observers. The detectability of calcifications, determined with a receiver operating characteristic (ROC) analysis area, was 0.92 for magnification and 0.82 for contact mammography with the single-emulsion-film system, compared with 0.84 and 0.72, respectively, with the dual-emulsion-film system. More clusters were correctly located and more individual calcifications were counted with magnification than with contact mammography. The dual-emulsion-film system with the magnification technique performs as well as the single-emulsion-film system with the contact technique, while retaining a decrease in required dose of approximately 40%.
在接触式和放大式乳腺摄影中,将一种新型双屏双乳剂胶片组合(可使辐射剂量降低约66%)与广泛使用的单屏单乳剂胶片系统进行了比较。对随机叠加在乳腺模型上的簇状微钙化进行检测,并由四名观察者确定单个钙化的位置和数量。通过接受者操作特征(ROC)分析区域确定钙化的可检测性,单乳剂胶片系统在放大式乳腺摄影中的可检测性为0.92,在接触式乳腺摄影中为0.82,而双乳剂胶片系统在放大式乳腺摄影和接触式乳腺摄影中的可检测性分别为0.84和0.72。与接触式乳腺摄影相比,放大式乳腺摄影能更准确地定位更多的簇状钙化,并能计数更多的单个钙化。双乳剂胶片系统结合放大技术的表现与单乳剂胶片系统结合接触技术的表现相当,同时所需剂量降低了约40%。