vd Drift J F, Brocaar M P, von Zanten G A, Lamoré P J
Department of Otorhinolaryngology, Dijkzigt University Hospital Rotterdam, The Netherlands.
Audiology. 1988;27(2):109-18. doi: 10.3109/00206098809081581.
In a test-retest experiment inaccuracies in the measurement of the peak latencies and threshold of the auditory brainstem response were determined for a group with normal hearing and for a group with cochlear hearing loss. The inaccuracy of the auditory brainstem response threshold is less than 4 dB in both groups. The inaccuracy in latency was measured as a function of stimulation level. In both groups the latency inaccuracy of peak V varies from 0.1 ms at levels well above threshold to 0.2 ms near the response threshold. Analysis of variance showed that in subjects with normal hearing the intra- and interindividual variabilities of the peak V latencies contribute about equally to the total variance at all stimulation levels. The implications that these findings have for the determination of the horizontal shift of the latency-intensity curve are discussed.
在一项重测实验中,针对听力正常组和耳蜗性听力损失组,测定了听觉脑干反应的峰值潜伏期和阈值测量中的误差。两组中听觉脑干反应阈值的误差均小于4分贝。潜伏期误差作为刺激水平的函数进行测量。在两组中,V波峰潜伏期误差在远高于阈值的水平时为0.1毫秒,在接近反应阈值时为0.2毫秒。方差分析表明,在听力正常的受试者中,V波峰潜伏期的个体内和个体间变异性在所有刺激水平下对总方差的贡献大致相等。讨论了这些发现对潜伏期-强度曲线水平偏移测定的影响。