Faculty of Applied Medical Sciences, Diagnostic Radiology, Jazan University, Jazan, Saudi Arabia.
School of Medicine, University of Nottingham, Nottingham, UK.
MAGMA. 2021 Dec;34(6):791-804. doi: 10.1007/s10334-021-00931-2. Epub 2021 Jun 5.
Quantitative Magnetic Resonance Imaging sequences have been investigated as objective imaging biomarkers of fibrosis and inflammation in Crohn's disease.
To determine the repeatability and inter- and intra-observer agreement of these measures in the prepared small bowel wall.
Ten healthy participants were scanned at 3 T on 2 separate occasions using T1 and T2 relaxometry, IVIM-DWI and MT sequences. Test-retest repeatability was assessed using the coefficient of variation (CoV) and intra-class correlation coefficients (ICCs) were used to evaluate the intra- and inter-observer agreement RESULTS: Test-retest repeatability in the bowel wall was excellent for apparent diffusion coefficient (ADC), magnetisation transfer ratio (MTR), T1, and diffusion coefficient D (CoV 5%, 7%, 8%, and 10%, respectively), good for perfusion fraction (PF) (CoV 20%) and acceptable for T2 (CoV 21%). Inter-observer agreement was good for the T2, D and ADC (ICC = 0.89, 0.86, 0.76, respectively) and moderate for T1 (ICC = 0.55). Intra-observer agreement was similar to inter-observer agreement.
This study showed variable results between the different parameters measured. Test-retest repeatability was at least acceptable for all parameters except pseudo-diffusion coefficient D*. Good inter- and intra-observer agreement was obtained for T2, ADC and D, with these parameters performing best in this technical validation study.
定量磁共振成像序列已被研究作为克罗恩病纤维化和炎症的客观成像生物标志物。
确定这些措施在准备好的小肠壁中的重复性和观察者内和观察者间的一致性。
10 名健康参与者在 3T 上分别在 2 次扫描中使用 T1 和 T2 弛豫度、IVIM-DWI 和 MT 序列进行扫描。使用变异系数 (CoV) 评估测试-重测重复性,使用组内相关系数 (ICC) 评估观察者内和观察者间的一致性。
肠壁的表观扩散系数 (ADC)、磁化传递率 (MTR)、T1 和扩散系数 D 的测试-重测重复性均为优秀(CoV 分别为 5%、7%、8%和 10%),灌注分数 (PF) 的重复性为良好(CoV 为 20%),T2 的重复性为可接受(CoV 为 21%)。T2、D 和 ADC 的观察者间一致性良好(ICC 分别为 0.89、0.86 和 0.76),T1 的一致性为中度(ICC 为 0.55)。观察者内一致性与观察者间一致性相似。
这项研究显示了所测量的不同参数之间的可变结果。除了伪扩散系数 D*外,所有参数的测试-重测重复性至少为可接受。T2、ADC 和 D 获得了良好的观察者内和观察者间一致性,这些参数在这项技术验证研究中表现最佳。