Greenspan M D, Lo C Y, Hanf D P, Yudkovitz J B
Department of Biochemical Regulation, Merck Institute for Therapeutic Research, Merck Sharp & Dohme Research Laboratories, Rahway, NJ 07065.
J Lipid Res. 1988 Jul;29(7):971-6.
A simple isocratic high performance liquid chromatography (HPLC) system is described that allows separation and identification of cholesteryl esters, triglycerides, ubiquinone, alpha-tocopherol, dolichol, cholesterol, 7-dehydrocholesterol, and retinol. This consisted of a normal phase cyanopropyl column with 0.1% isopropanol in heptane as the solvent. The effluent was monitored with an LKB model 2140 diode array detector which enabled the lipids to be identified by their characteristic absorption spectra. This system was applied to a sample of dog liver in which cholesteryl esters, retinyl esters, triglycerides, ubiquinone, dolichol, cholesterol, and retinol were identified. Retinyl esters and vitamin D esters were identified by their similarity in absorption spectra to retinol and vitamin D. A system to transfer and store the chromatograms on the VAX PDP-11 or an optical disc is also described.
本文描述了一种简单的等度高效液相色谱(HPLC)系统,该系统可用于分离和鉴定胆固醇酯、甘油三酯、泛醌、α-生育酚、多萜醇、胆固醇、7-脱氢胆固醇和视黄醇。该系统由一个正相氰丙基柱组成,以庚烷中0.1%的异丙醇作为溶剂。流出物用LKB 2140型二极管阵列检测器进行监测,该检测器能够根据脂质的特征吸收光谱对其进行鉴定。该系统应用于犬肝样本,从中鉴定出了胆固醇酯、视黄醇酯、甘油三酯、泛醌、多萜醇、胆固醇和视黄醇。视黄醇酯和维生素D酯通过其吸收光谱与视黄醇和维生素D的相似性得以鉴定。本文还介绍了一种在VAX PDP - 11或光盘上传输和存储色谱图的系统。