基于显色离子载体的血清和血浆中钠和钾的分光光度测定方法。

Chromogenic ionophore-based methods for spectrophotometric assay of sodium and potassium in serum and plasma.

作者信息

Kumar A, Chapoteau E, Czech B P, Gebauer C R, Chimenti M Z, Raimondo O

机构信息

Technicon Instruments Corp., Tarrytown, NY 10591.

出版信息

Clin Chem. 1988 Sep;34(9):1709-12.

PMID:3416420
Abstract

We report spectrophotometric methods for quantifying sodium and potassium in serum and plasma without sample pretreatment or solvent-extraction steps. The methods are based on novel chromogenic ionophores characterized by molecular structures highly preorganized for binding, which result in exceptionally high sensitivity and selectivity (negligible responses to sodium at 80 to 200 mmol/L in the potassium assay and to potassium at 0 to 10 mmol/L in the sodium assay). The within-run imprecisions are comparable with those for ion-selective electrodes (for K, 1.0% at 4.5 mmol/L; for Na, 1.1% at 136 mmol/L); the analytical range is 2-10 mmol/L for potassium, 80-170 mmol/L for sodium. The results for greater than 100 patients' samples correlated well with results obtained with ion-selective electrodes (r = 0.99 for K, 0.97 for Na). The ready-to-use aqueous reagents will be available under the trade name of ChromoLyte for use with Technicon's RA family of clinical analyzers.

摘要

我们报告了无需样品预处理或溶剂萃取步骤即可定量血清和血浆中钠和钾的分光光度法。这些方法基于新型显色离子载体,其特征在于具有高度预组织化的用于结合的分子结构,这导致了极高的灵敏度和选择性(在钾测定中,80至200 mmol/L的钠对测定结果影响可忽略不计;在钠测定中,0至10 mmol/L的钾对测定结果影响可忽略不计)。批内不精密度与离子选择性电极相当(对于钾,在4.5 mmol/L时为1.0%;对于钠,在136 mmol/L时为1.1%);钾的分析范围为2至10 mmol/L,钠的分析范围为80至170 mmol/L。100多名患者样本的结果与离子选择性电极获得的结果相关性良好(钾的r = 0.99,钠的r = 0.97)。即用型水性试剂将以ChromoLyte的商品名供应,用于Technicon的RA系列临床分析仪。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验