Stability of reacted Chemstrip bG.

作者信息

Cox D, Herrman J, Snyder A, Gonder-Frederick L, Reschke J, Clarke W

机构信息

Department of Behavioral Medicine, University of Virginia Medical School, Charlottesville 22901.

出版信息

Diabetes Care. 1988 Mar;11(3):288-91. doi: 10.2337/diacare.11.3.288.

Abstract

Reacted Chemstrip bG glucose reagent strips have been reported to retain their color changes for up to 7 days. Thus, patients could theoretically measure their blood glucose and mail their reacted test strips to their physicians for reanalysis. To test the stability of reacted Chemstrip bG blood glucose measurements, 268 Chemstrip bG test strips were reacted with blood obtained from 67 insulin-dependent diabetic patients, stored in desiccator vials, and read daily for 5 consecutive days with an Accu-Chek II blood glucose meter. Although Chemstrip bG blood glucose values significantly correlated with initial reference Beckman glucose analyzer glucose determinations for all 5 days, a steady significant decay in blood glucose readings over time was observed, and clinically accurate strip readings declined from 94% to 68%. Because this decay appeared consistent, correction factors were calculated with regression analyses. The correction factor for day 5 Accu-Chek II readings reduced measurement error by 77%. When applied to a different validation sample, this correction factor decreased day 5 error by 73%. Hence, it seems that correction factors may be applied to delayed readings of Chemstrips obtained with Accu-Chek II that would correct for the observed reduction in blood glucose readings. From these results, we conclude that delayed readings of Chemstrip bG test strips with the Accu-Chek II are not sufficiently accurate for clinical decision making or research purposes unless mathematically corrected.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验