[用于Selectron环形施源器系统(高剂量率和低剂量率后装)的直肠屏蔽]

[Rectal shielding for a Selectron-ring applicator system (HDR- and LDR-afterloading)].

作者信息

Hetzel H, Kamleitner H, McCoy M, Frommhold H

机构信息

Universitätsklinik für Frauenheilkunde, Innsbruck.

出版信息

Strahlenther Onkol. 1987 Dec;163(12):782-6.

PMID:3424077
Abstract

In order to decrease the morbidity rate after combined radiotherapy of the cervix carcinoma, a tungsten shield 3 and 5 mm thick for the rectum has been developed by the authors which is applied with the ring and pin applicator of the Selectron unit (LDR- and HDR-afterloading). The isodose curves were measured in a plexiglas phantom, and the radiation dose at the reference points was determined by means of a ionization dosimeter. The phantom measurements were performed with the same arrangement of sources as applied in radiotherapy. The measurements showed a dose reduction at point Rmax of 33% (HDR) and 44% (LDR) with the tungsten shield 5 mm thick.

摘要

为降低宫颈癌联合放疗后的发病率,作者研制了一种3毫米和5毫米厚的用于直肠的钨屏蔽物,该屏蔽物与Selectron装置(近距离后装放疗的低剂量率和高剂量率)的环形和针形施源器一起使用。在有机玻璃模型中测量了等剂量曲线,并通过电离剂量计确定参考点处的辐射剂量。模型测量采用与放疗中相同的源排列方式进行。测量结果显示,使用5毫米厚的钨屏蔽物时,Rmax点的剂量降低了33%(高剂量率)和44%(低剂量率)。