Meertens H, van der Laarse R
Radiother Oncol. 1985 Jan;3(1):69-80. doi: 10.1016/s0167-8140(85)80010-4.
The addition of screens in the vaginal source holders of a cervix applicator for intracavitary brachytherapy reduces the dose to rectum and bladder and therefore diminishes the number of rectal and vesical complications. Shielding properties of tungsten rectal and bladder screens of a Selectron cervix applicator, loaded with spherical cesium sources, were determined for verification of dose calculations. Transmission characteristics of half-disk shaped tungsten screen segments in a single ovoid were measured in a water phantom. The minimum transmission ratios are 60, 70 and 80% for segment thickness of 5.0, 3.5 and 2.0 mm, respectively. The accuracy of the new screen correction algorithm of the Selectron Planning System was assessed by comparing measured and calculated dose rates and was found to be better than +/- 4%. The correction algorithm provides a method to analyse the efficacy of screens in the ovoids for various segment geometries and orientations without extensive phantom measurements. Isotransmission and isodose calculations were made for a typical clinical applicator set-up and source distribution. The dose reduction to rectum and bladder, near the bottom and top of the ovoids was analysed in detail. A 3.5 mm thick rectum and bladder screen in each ovoid reduces the dose approximately by 20% to the rectum and by 15% to the bladder. A distance enlargement of about 5 mm between ovoid and rectum or bladder, e.g. by packing, results in a comparable dose reduction. Shielding properties of a Selectron cervix applicator, provided with screens, were compared with those of some Fletcher-type applicators. Significant differences between the transmission ratios and shielded areas of the screens of both systems near rectum and bladder were observed.