Suppr超能文献

计算算法和半影:剂量学审核中危及器官剂量低估。

Calculation algorithms and penumbra: Underestimation of dose in organs at risk in dosimetry audits.

机构信息

Australian Clinical Dosimetry Service, ARPANSA, Yallambie, Victoria, Australia.

Physical Sciences, Peter MacCallum Cancer Centre, Melbourne, Victoria, Australia.

出版信息

Med Phys. 2021 Oct;48(10):6184-6197. doi: 10.1002/mp.15123. Epub 2021 Aug 18.

Abstract

PURPOSE

The aim of this study is to investigate overdose to organs at risk (OARs) observed in dosimetry audits in Monte Carlo (MC) algorithms and Linear Boltzmann Transport Equation (LBTE) algorithms. The impact of penumbra modeling on OAR dose was assessed with the adjustment of MC modeling parameters and the clinical relevance of the audit cases was explored with a planning study of spine and head and neck (H&N) patient cases.

METHODS

Dosimetric audits performed by the Australian Clinical Dosimetry Service (ACDS) of 43 anthropomorphic spine plans and 1318 C-shaped target plans compared the planned dose to doses measured with ion chamber, microdiamond, film, and ion chamber array. An MC EGSnrc model was created to simulate the C-shape target case. The electron cut-off energy E was set at 500, 200, and 10 keV, and differences between 1 and 3 mm voxel were calculated. A planning study with 10 patient stereotactic body radiotherapy (SBRT) spine plans and 10 patient H&N plans was calculated in both Acuros XB (AXB) v15.6.06 and Anisotropic Analytical Algorithm (AAA) v15.6.06. The patient contour was overridden to water as only the penumbral differences between the two different algorithms were under investigation.

RESULTS

The dosimetry audit results show that for the SBRT spine case, plans calculated in AXB are colder than what is measured in the spinal cord by 5%-10%. This was also observed for other audit cases where a C-shape target is wrapped around an OAR where the plans were colder by 3%-10%. Plans calculated with Monaco MC were colder than measurements by approximately 7% with the OAR surround by a C-shape target, but these differences were not noted in the SBRT spine case. Results from the clinical patient plans showed that the AXB was on average 7.4% colder than AAA when comparing the minimum dose in the spinal cord OAR. This average difference between AXB and AAA reduced to 4.5% when using the more clinically relevant metric of maximum dose in the spinal cord. For the H&N plans, AXB was cooler on average than AAA in the spinal cord OAR (1.1%), left parotid (1.7%), and right parotid (2.3%). The EGSnrc investigation also noted similar, but smaller differences. The beam penumbra modeled by E  = 500 keV was steeper than the beam penumbra modeled by E  = 10 keV as the full scatter is not accounted for, which resulted in less dose being calculated in a central OAR region where the penumbra contributes much of the dose. The dose difference when using 2.5 mm voxels of the center of the OAR between 500 and 10 keV was 3%, reducing to 1% between 200 and 10 keV.

CONCLUSIONS

Lack of full penumbral modeling due to approximations in the algorithms in MC based or LBTE algorithms are a contributing factor as to why these algorithms under-predict the dose to OAR when the treatment volume is wrapped around the OAR. The penumbra modeling approximations also contribute to AXB plans predicting colder doses than AAA in areas that are in the vicinity of beam penumbra. This effect is magnified in regions where there are many beam penumbras, for example in the spinal cord for spine SBRT cases.

摘要

目的

本研究旨在调查蒙特卡罗(MC)算法和线性 Boltzmann 输运方程(LBTE)算法中的剂量学审计中观察到的危及器官(OAR)的剂量超过。通过调整 MC 建模参数评估了半影建模对 OAR 剂量的影响,并通过脊柱和头颈部(H&N)患者病例的计划研究探讨了审计病例的临床相关性。

方法

澳大利亚临床剂量服务(ACDS)对 43 个人体脊柱计划和 1318 个 C 形靶计划进行了剂量学审计,将计划剂量与电离室、微钻、胶片和电离室阵列测量的剂量进行了比较。创建了一个 MC EGSnrc 模型来模拟 C 形靶计划。电子截止能量 E 设置为 500keV、200keV 和 10keV,并计算了 1mm 和 3mm 体素之间的差异。在 Acuros XB(AXB)v15.6.06 和各向异性解析算法(AAA)v15.6.06 中分别对 10 例立体定向体部放射治疗(SBRT)脊柱计划和 10 例 H&N 计划进行了计划研究。仅研究两种不同算法之间的半影差异,因此覆盖了患者轮廓到水。

结果

剂量学审计结果表明,对于 SBRT 脊柱病例,AXB 计算的计划比脊髓中测量的温度低 5%-10%。在其他审计病例中,当 C 形靶标包裹在 OAR 周围时,也观察到了类似的情况,计划的温度比 OAR 低 3%-10%。使用 Monaco MC 计算的计划比测量值低约 7%,当 OAR 周围有 C 形靶标时,但在 SBRT 脊柱病例中未注意到这些差异。来自临床患者计划的结果表明,当比较脊髓 OAR 的最小剂量时,AXB 平均比 AAA 低 7.4%。当使用脊髓 OAR 中的最大剂量更具临床相关性的指标进行比较时,AXB 和 AAA 之间的平均差异降低到 4.5%。对于 H&N 计划,AXB 在脊髓 OAR(1.1%)、左腮腺(1.7%)和右腮腺(2.3%)中比 AAA 平均更冷。EGSnrc 调查还注意到类似但较小的差异。E = 500keV 建模的束半影比 E = 10keV 建模的束半影更陡峭,因为没有考虑完全散射,这导致在束半影贡献大部分剂量的中央 OAR 区域中计算的剂量减少。在 500keV 和 10keV 之间,OAR 中心 2.5mm 体素的剂量差异为 3%,在 200keV 和 10keV 之间减少到 1%。

结论

由于 MC 或 LBTE 算法中的算法近似值缺乏完整的半影建模,这是这些算法在治疗体积包裹 OAR 时低估 OAR 剂量的原因之一。半影建模的近似值也导致 AXB 计划预测的剂量比 AAA 更冷,在束半影附近的区域。在有许多束半影的区域,例如脊柱 SBRT 病例中的脊髓,这种影响会放大。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验