Department of Otorhinolaryngology-Head and Neck Surgery, Seoul National University Hospital, Seoul, South Korea.
Department of Otorhinolaryngology-Head and Neck Surgery, Seoul National University Bundang Hospital, Seongnam, South Korea.
Sci Rep. 2021 Jul 23;11(1):15135. doi: 10.1038/s41598-021-94668-6.
The slim modiolar electrode has been reported to ensure better modiolar proximity than previous conventional perimodiolar electrodes and consistently high scala tympani localization. Nonetheless, variability in modiolar proximity exists even among slim modiolar electrodes, still leaving room for further improvement of modiolar proximity, which may positively affect functional outcomes. Given this, the pull-back maneuver was reported to increase the modiolar proximity of slim modiolar electrodes in a cadaveric study, but in vivo repositioning effects remain to be established. Here we identified that the pull-back maneuver led to better modiolar proximity than conventional insertion while maintaining a similar angular insertion depth. Notably, the reduced electrode-modiolus distance from the pull-back maneuver was associated with significantly lower impedances across electrodes postoperatively as well as reduced intraoperative electrophysiological thresholds than conventional insertion. Among adult cochlear implant recipients, this maneuver resulted in significantly better sentence recognition scores at three months postoperatively when compared to those with a conventional insertion; however, this benefit was not observed at later intervals. Collectively, slim modiolar electrodes with the pull-back maneuver further enhance the modiolar proximity, possibly leading to better open-set sentence recognition, at least in the early postoperative stage.
薄的圆窗电极已被报道能比以前的传统的圆窗内电极更好地确保接近圆窗,并且能始终保持较高的鼓阶定位。尽管如此,即使是在薄的圆窗电极中,也存在圆窗接近度的可变性,这仍然为进一步改善圆窗接近度留下了空间,而这可能对功能结果产生积极影响。考虑到这一点,在一项尸体研究中报道了后退操作可增加薄的圆窗电极的圆窗接近度,但体内重新定位效果仍有待确定。在这里,我们发现后退操作比传统插入能更好地接近圆窗,同时保持类似的角插入深度。值得注意的是,与传统插入相比,从后退操作获得的电极-圆窗距离减小与术后电极之间的阻抗显著降低以及术中电生理阈值降低有关。在成年人工耳蜗植入者中,与传统插入相比,该操作在术后三个月时导致明显更好的句子识别分数;然而,在后期间隔时未观察到这种益处。总体而言,带有后退操作的薄的圆窗电极进一步增强了圆窗接近度,可能导致更好的开放式句子识别,至少在术后早期阶段是这样。