Suppr超能文献

利用智能手机管理介入放射学中的职业剂量。

Managing occupational doses with smartphones in interventional radiology.

机构信息

Medical Physics, Hospital Clínico San Carlos, Madrid, Spain.

Instituto de Investigación Sanitaria del Hospital Clínico San Carlos, Madrid, Spain.

出版信息

Med Phys. 2021 Oct;48(10):5830-5836. doi: 10.1002/mp.15148. Epub 2021 Aug 17.

Abstract

PURPOSE

This study presents a prototype smartphone application for occupational dosimetry in interventional practices based on electronic personal dosimeters to assist in dose monitoring.

METHODS

The prototype receives and records information from the occupational dose report containing the cumulative dose of electronic personal dosimeters worn over the apron at chest level and electronic area dosimeters located on C-arms (reference dosimeters), for each fluoroscopy-guided procedure. Using their smartphones, personnel involved in interventional practices can review and compare their occupational records with an investigation level, the dose limits, and their department colleagues (anonymously). The ratio between H (10) measured by the personal and the reference dosimeters at the C-arm is presented as an indicator of consistent use of suspended operator shield. Some general results extracted from the first months of use are presented.

RESULTS

The reference dosimeter located at the C-arm (without lead protection and acting as an ambient dosimeter) recorded in one of the laboratories 217 mSv during 308 procedures over 5 months, showing an indication of the radiation risk present in an interventional laboratory. The ratio between the personal cumulative dose and the dose at a reference C-arm dosimeter ranged from 0.2% to 1.67% (a factor of 8.5) for different interventionalists. These differences suggest different protection habits among interventional operators, as well as a target for dose reduction.

CONCLUSIONS

With this system, professionals have easy access to their occupational dosimetry records (including information on the workload) in the setting of their interventional departments, to thereby actively engage in the protection process.

摘要

目的

本研究提出了一种基于电子个人剂量计的介入实践职业剂量原型智能手机应用程序,以协助剂量监测。

方法

原型机接收并记录来自职业剂量报告的信息,该报告包含佩戴在胸前围裙上的电子个人剂量计和 C 臂上的电子区域剂量计(参考剂量计)的累积剂量,用于每次透视引导的程序。介入实践人员可以使用他们的智能手机查看和比较他们的职业记录与调查水平、剂量限制以及他们部门的同事(匿名)。个人剂量计和 C 臂上的参考剂量计测量的 H(10)比值作为一致使用悬挂式操作人员屏蔽的指示。展示了从使用的第一个月提取的一些一般结果。

结果

位于 C 臂的参考剂量计(没有铅保护,充当环境剂量计)在五个月内记录了一个实验室中的 217 mSv,在 308 次手术中,表明介入实验室存在辐射风险。个人累积剂量与参考 C 臂剂量计剂量之间的比值在不同的介入医生之间从 0.2%到 1.67%(8.5 倍)不等。这些差异表明介入操作人员之间存在不同的防护习惯,以及降低剂量的目标。

结论

通过该系统,专业人员可以在介入科室环境中轻松访问他们的职业剂量记录(包括工作量信息),从而积极参与防护过程。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验