Gehrke Thorsten, Citak Mustafa, Abdelaziz Hussein
Department of Orthopaedic Surgery, Helios ENDO-Klinik Hamburg, Hamburg, Germany.
Arthroplast Today. 2021 Jul 19;10:108-113. doi: 10.1016/j.artd.2021.06.011. eCollection 2021 Aug.
Catastrophic failure of ceramic-on-polyethylene total hip arthroplasty is still occasionally described. We report on a new case of complete atraumatic penetration of an intact ceramic head through the titanium cup in a cementless total hip arthroplasty due to dramatic polyethylene and metal wear. We reviewed the literature for similar cases and analyzed potential risk factors. Most importantly, revision of radiologically worn liners should not be postponed, especially in young and active patients with conventional liners, because the time to dramatic failure could be shorter than expected.
陶瓷对聚乙烯全髋关节置换术的灾难性失败仍时有报道。我们报告了一例新病例,在无水泥全髋关节置换术中,由于聚乙烯和金属的剧烈磨损,完整的陶瓷股骨头无创伤地穿透钛杯。我们回顾了文献中的类似病例并分析了潜在风险因素。最重要的是,对于影像学显示磨损的内衬,不应推迟翻修,尤其是对于使用传统内衬的年轻活跃患者,因为发生严重失败的时间可能比预期的要短。