Clin Oral Implants Res. 2021 Oct;32(10):1190-1199. doi: 10.1111/clr.13816. Epub 2021 Aug 14.
To analyze the early stages of osseointegration around implants with different geometry following installation in sites using different osteotomy protocols.
Two types of implants were installed using regular or modified (reduced diameter) osteotomy protocols in mandibular premolar/molar regions following tooth extraction in six dogs. Three implant site categories were created: Reference (A implant and regular osteotomy), Test-1 (B implant and regular osteotomy), and Test-2 (B implant and modified osteotomy). Implant installation procedures were repeated after 4 and 6 weeks. The insertion torque (ITQ) was measured during implant installation and resonance frequency analysis providing implant stability quotient (ISQ) values was performed following implant installation and once every week during the course of the study. Biopsies were obtained immediately after the 3rd installation procedure and prepared for histological analysis.
The modified osteotomy protocol created a higher insertion torque (ITQ). The analysis also revealed a correlation between insertion torque at implant installation and radiographic bone loss after 6 weeks of healing. The decline in ISQ values during healing was more pronounced at Test-2 than other sites. While the degree of bone-to-implant contact (BIC%) in the marginal area was similar in test and reference sites at 6 weeks of healing, Test-2 implants presented with a significantly higher BIC% in the mid and apical areas than Test-1 implants.
It is suggested that placement of implants in undersized osteotomy sites will result in an increased remodeling of the cortical bone during the early healing process.
分析不同几何形状的种植体在不同骨切开术方案的种植部位早期骨整合的情况。
在 6 只狗拔牙后的下颌前磨牙/磨牙区,使用常规或改良(减小直径)骨切开术方案安装两种类型的种植体。创建了三种种植体部位类别:参考(A 种植体和常规骨切开术)、测试-1(B 种植体和常规骨切开术)和测试-2(B 种植体和改良骨切开术)。在 4 周和 6 周后重复种植体安装程序。在种植体安装过程中测量插入扭矩(ITQ),并在种植体安装后和研究过程中的每周一次进行共振频率分析以提供种植体稳定性指数(ISQ)值。在第 3 次安装程序后立即获得活检并进行组织学分析。
改良骨切开术方案产生了更高的插入扭矩(ITQ)。分析还显示,种植体安装时的插入扭矩与 6 周愈合后的放射影像学骨丢失之间存在相关性。在测试-2 中,ISQ 值在愈合过程中的下降比其他部位更为明显。虽然在 6 周愈合时,测试和参考部位的边缘区域的骨-种植体接触度(BIC%)相似,但测试-2 种植体的中、根尖区域的 BIC%明显高于测试-1 种植体。
建议在尺寸不足的骨切开部位放置种植体,将导致在早期愈合过程中皮质骨的重塑增加。