斯洛文尼亚典型成人 CT 剂量指标在 PET-CT 扫描中的建立。

Establishment of typical adult CT dose indicators for PET-CT scans in Slovenia.

机构信息

Institute of Oncology Ljubljana, Nuclear medicine department, Zaloška cesta 2, Ljubljana, Slovenia.

University of Ljubljana, Faculty of Health Sciences, Medical Imaging and Radiotherapy Department, Zdravstvena pot 5, Ljubljana, Slovenia.

出版信息

J Radiol Prot. 2021 Aug 19;41(3). doi: 10.1088/1361-6498/ac025a.

Abstract

The aim was to determine typical values of diagnostic reference level (DRL) quantities for the computed tomography (CT) part of the most common positron emission tomography-computed tomography (PET-CT) procedures in Slovenia. The most common PET-CT procedures were identified, and data collated for 565 patients imaged in all three PET-CT units in Slovenia during a time span of 11 months. As the number of facilities is too low to establish national DRLs, we followed ICRP recommendations and determined typical values of DRL quantities as the median values of the pooled set of data. Mean, median, and standard deviation of CT dose index (CTDI) and total dose length product (DLP) for the CT part of the most common PET-CT procedures were determined for pooled data as well as for each PET-CT unit. The data were compared between all three units to identify possible outliers that would likely benefit from optimization. Three most common CT protocols covering approximately 2/3 of all PET-CT imaging performed in Slovenia were considered: from the base of the cranium to the middle of the femur, from the top of the head to the middle of the femur, and for the whole-body PET-CT. The established typical values in terms of total DLP were 295, 359, and 676 mGyċcm, respectively; and in terms of CTDI3.05, 3.22, and 3.60 mGy, respectively. Comparing the data between all three units showed significantly higher (< 0.001) patient doses on one unit, indicating a need for optimization. The results present the first-time data on the national typical values of DRL quantities for the CT part of most common PET-CT procedures in Slovenia. While the determined typical values are within the DRL values established in some other countries, significant differences were found between the individual units included in the study.

摘要

目的是确定斯洛文尼亚最常见的正电子发射断层扫描-计算机断层扫描(PET-CT)程序中 CT 部分的诊断参考水平(DRL)量的典型值。确定了最常见的 PET-CT 程序,并在 11 个月的时间内对斯洛文尼亚所有三个 PET-CT 单位的 565 名患者进行了成像,收集了数据。由于设施数量太少,无法制定国家 DRL,因此我们遵循 ICRP 的建议,将 DRL 量的典型值确定为数据汇总集的中位数。还为最常见的 PET-CT 程序的 CT 部分确定了 CT 剂量指数(CTDI)和总剂量长度乘积(DLP)的平均值、中位数和标准差,以及汇总数据和每个 PET-CT 单位的数据。比较了所有三个单位之间的数据,以确定可能受益于优化的异常值。考虑了覆盖斯洛文尼亚大约 2/3 的所有 PET-CT 成像的三个最常见的 CT 协议:从颅底到股骨中部、从头顶到股骨中部以及全身 PET-CT。以总 DLP 表示的建立的典型值分别为 295、359 和 676 mGyċcm;以 CTDI3.05 表示的典型值分别为 3.22 和 3.60 mGy。比较所有三个单位的数据显示,一个单位的患者剂量明显更高(<0.001),表明需要优化。这些结果提供了斯洛文尼亚最常见的 PET-CT 程序中 CT 部分的 DRL 量的全国典型值的首次数据。虽然确定的典型值在一些其他国家建立的 DRL 值范围内,但在所研究的个别单位之间发现了显著差异。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验