Intravoxel incoherent motion (IVIM) studies are performed with different acquisition protocols. Comparing them requires knowledge of echo time (TE) dependencies. The TE-dependence of the biexponential perfusion fraction f is well-documented, unlike that of its triexponential counterparts f and f and the biexponential and triexponential pseudodiffusion coefficients D , , and . The purpose was to investigate the TE-dependence of these parameters and to check whether the triexponential pseudodiffusion compartments are associated with arterial and venous blood.
Fifteen healthy volunteers (19-58 y; mean: 24.7 y) underwent diffusion-weighted imaging of the abdomen with 24 b-values (0.2-800 s/mm ) at TEs of 45, 60, 75, and 90 ms. Regions of interest (ROIs) were manually drawn in the liver. One set of bi- and triexponential IVIM parameters per volunteer and TE was determined. The TE-dependence was assessed with the Kruskal-Wallis test.
TE-dependence was observed for f (P < .001), f (P = .001), and f (P < .001). Their median values at the four measured TEs were: f: 0.198/0.240/0.274/0.359, f : 0.113/0.139/0.146/0.205, f : 0.115/0.155/0.182/0.194. D, D , , and showed no significant TE-dependence. Their values were: diffusion coefficient D (10 mm /s): 9.45/9.63/9.75/9.41, biexponential D (10 mm /s): 5.26/5.52/6.13/5.82, triexponential (10 mm /s): 1.73/2.91/2.25/2.51, triexponential (mm /s): 0.478/1.385/0.616/0.846.
f and f show similar TE-dependence as f, ie, increase with rising TE; an effect that must be accounted for when comparing different studies. The diffusion and pseudodiffusion coefficients might be compared without TE correction. Because of the similar TE-dependence of f and f , the triexponential pseudodiffusion compartments are most probably not associated to venous and arterial blood.
体素内不相干运动(IVIM)研究采用不同的采集方案。比较这些方案需要了解回波时间(TE)依赖性。双指数灌注分数f的TE依赖性已有充分记录,与其三指数对应物f和f以及双指数和三指数伪扩散系数D、 、 不同。目的是研究这些参数的TE依赖性,并检查三指数伪扩散区是否与动脉血和静脉血相关。
15名健康志愿者(19 - 58岁;平均:24.7岁)在45、60、75和90 ms的TE下,使用24个b值(0.2 - 800 s/mm )对腹部进行扩散加权成像。在肝脏中手动绘制感兴趣区域(ROI)。确定每位志愿者和每个TE的一组双指数和三指数IVIM参数。使用Kruskal - Wallis检验评估TE依赖性。
观察到f(P <.001)、f(P =.001)和f(P <.001)存在TE依赖性。它们在四个测量TE下的中位数分别为:f:0.198/0.240/0.274/0.359,f :0.113/0.139/0.146/0.205,f :0.115/0.155/0.182/0.194。D、D 、 和 未显示出显著的TE依赖性。它们的值分别为:扩散系数D(10 mm /s):9.45/9.63/9.75/9.41,双指数D(10 mm /s):5.26/5.52/6.13/5.82,三指数 (10 mm /s):1.73/2.91/2.25/2.51,三指数 (mm /s):0.478/1.385/0.616/0.846。
f和f与f表现出相似的TE依赖性,即随TE升高而增加;在比较不同研究时必须考虑这一效应。扩散系数和伪扩散系数在不进行TE校正的情况下也可比较。由于f和f具有相似的TE依赖性,三指数伪扩散区很可能与静脉血和动脉血无关。