Oh Se An, Kim Sung Yeop, Park Jaehyeon, Park Jae Won, Yea Ji Woon
Department of Radiation Oncology, Yeungnam University College of Medicine, Daegu, Korea.
Department of Physics, Yeungnam University, Gyeongsan, Korea.
J Yeungnam Med Sci. 2022 Apr;39(2):108-115. doi: 10.12701/yujm.2021.01123. Epub 2021 Sep 9.
This study was aimed at comparing and analyzing the results of FractionLab (Varian/Mobius Medical System) with those of portal dosimetry that uses an electronic portal imaging device. Portal dosimetry is extensively used for patient-specific quality assurance (QA) in intensity-modulated radiotherapy (IMRT).
The study includes 29 patients who underwent IMRT on a Novalis-Tx linear accelerator (Varian Medical System and BrainLAB) between June 2019 and March 2021. We analyzed the multileaf collimator (MLC) DynaLog files generated after portal dosimetry to evaluate the same condition using FractionLab. The results of the recently launched FractionLab at various gamma indices (0.1%/0.1 mm-1%/1 mm) are analyzed and compared with those of portal dosimetry (3%/3 mm).
The average gamma passing rates of portal dosimetry (3%/3 mm) and FractionLab are 98.1 (95.5%-100%) and 97.5% (92.3%-99.7%) at 0.6%/0.6 mm, respectively. The results of portal dosimetry (3%/3 mm) are statistically comparable with the QA results of FractionLab (0.6%/0.6 mm-0.9%/0.9 mm).
This paper presents the clinical performance of FractionLab by the comparison of the QA results of FractionLab using portal dosimetry with various gamma indexes when performing patient-specific QA in IMRT treatment. Further, the appropriate gamma index when performing patient-specific QA with FractionLab is provided.
本研究旨在比较和分析FractionLab(瓦里安/莫比乌斯医疗系统)与使用电子射野影像装置的射野剂量测定法的结果。射野剂量测定法广泛用于调强放射治疗(IMRT)中针对患者的质量保证(QA)。
该研究纳入了29例在2019年6月至2021年3月期间在诺瓦利斯-Tx直线加速器(瓦里安医疗系统和BrainLAB)上接受IMRT治疗的患者。我们分析了射野剂量测定后生成的多叶准直器(MLC)动态日志文件,以使用FractionLab评估相同情况。分析了最近推出的FractionLab在各种伽马指数(0.1%/0.1毫米 - 1%/1毫米)下的结果,并与射野剂量测定法(3%/3毫米)的结果进行比较。
在0.6%/0.6毫米时,射野剂量测定法(3%/3毫米)和FractionLab的平均伽马通过率分别为98.1(95.5% - 100%)和97.5%(92.3% - 99.7%)。射野剂量测定法(3%/3毫米)的结果与FractionLab(0.6%/0.6毫米 - 0.9%/0.9毫米)的QA结果在统计学上具有可比性。
本文通过在IMRT治疗中进行针对患者的QA时,比较使用射野剂量测定法的FractionLab在各种伽马指数下的QA结果,展示了FractionLab的临床性能。此外,还提供了使用FractionLab进行针对患者的QA时合适的伽马指数。