手持测力计评估肩部旋转肌力量的绝对可靠性和同时效度:系统评价和荟萃分析。

Absolute Reliability and Concurrent Validity of Hand-Held Dynamometry in Shoulder Rotator Strength Assessment: Systematic Review and Meta-Analysis.

机构信息

Faculty of Rehabilitation Sciences, Universidad Andres Bello, Santiago 7591538, Chile.

Servicio Kinesioterapia Ambulatoria, Clínica Las Condes, Santiago 7591538, Chile.

出版信息

Int J Environ Res Public Health. 2021 Sep 3;18(17):9293. doi: 10.3390/ijerph18179293.

Abstract

The purpose of this study is to establish the absolute reliability between hand-held dynamometers (HHDs) and concurrent validity between HHDs and isokinetic dynamometers (IDs) in shoulder rotator strength assessment. The Medline, CINAHL, and Central databases were searched for relevant studies up to July 2020. Absolute reliability was determined by test-retest studies presenting standard error of measurement (SEM%) and/or minimal detectable change (MDC%) expressed as percentage of the mean. Studies considering intra-class correlation coefficient (ICC) between IDs and HHDs were considered for concurrent validity. The risk of bias and the methodological quality were evaluated according to COnsensus-based Standards for the selection of health Measurement INstruments (COSMIN). Thirteen studies were included in the meta-analysis. Shoulder internal rotator strength assessment MDC% was 0.78%, 95% confidence interval (CI) -5.21 to 3.66, while shoulder external rotators MDC% was 3.29%, CI -2.69 to 9.27. ICC between devices was 0.94, CI (0.91 to 0.96) for shoulder internal rotators and 0.92, IC (0.88 to 0.97) for shoulder external rotators. Very high correlation was found for shoulder rotator torque assessment between HHDs and IDs. The COSMIN checklist classified the selected studies as adequate and inadequate.

摘要

本研究旨在确定手持式测力计(HHD)之间的绝对可靠性,以及 HHD 与等速测力计(ID)之间在肩部旋转力量评估中的同时效度。检索了截至 2020 年 7 月的 Medline、CINAHL 和中央数据库中的相关研究。绝对可靠性通过呈现测量标准误差(SEM%)和/或最小可检测变化(MDC%)的测试-重测研究来确定,以平均值的百分比表示。考虑 ID 和 HHD 之间的组内相关系数(ICC)的研究被认为具有同时效度。根据共识基础健康测量仪器选择标准(COSMIN)评估偏倚风险和方法学质量。共有 13 项研究纳入荟萃分析。肩部内旋力量评估的 MDC%为 0.78%,95%置信区间(CI)为-5.21 至 3.66,而肩部外旋力量评估的 MDC%为 3.29%,CI 为-2.69 至 9.27。设备之间的 ICC 为 0.94,CI(0.91 至 0.96)用于肩部内旋,0.92,IC(0.88 至 0.97)用于肩部外旋。HHD 和 ID 之间对肩部旋转力量评估的相关性非常高。COSMIN 清单将所选研究归类为充分和不充分。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/52c3/8430567/2848d46d65d0/ijerph-18-09293-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验