Evaluation of a novel method to calculate cementless femoral stem level on craniocaudal projection radiographs.

机构信息

Animal Emergency & Specialty Center, Parker, Colorado, USA.

Nexus Veterinary Specialists, Victoria, Texas, USA.

出版信息

Vet Surg. 2021 Nov;50(8):1592-1599. doi: 10.1111/vsu.13723. Epub 2021 Sep 20.

Abstract

OBJECTIVE

To assess the accuracy of a proportion-corrected radiographic cementless femoral stem level calculation in a cadaver model.

STUDY DESIGN

Cadaveric study.

SAMPLE POPULATION

Twelve cadaveric canine femurs.

METHODS

Cementless femoral stems were implanted in 12 cadaver femurs at three stem levels. Craniocaudal projection radiographs of the implanted femurs were obtained at 0°, 10°, 20°, 30°, and 45° angulations in the sagittal plane (relative to the radiographic plate) and at 0°, 15°, and 30° internal and external axial rotation. A proportional calculation including actual stem length, radiographically measured stem length, and radiographically measured stem level relative to the greater trochanter (uncorrected stem level) was solved to yield a corrected stem level. The errors of the stem level measurements were compared.

RESULTS

With 0° or 15° of internal rotation, the true stem level differed by more than 0.5 mm from corrected stem level in 2 of 30 tested femoral positions. It differed from the uncorrected stem level in 10 of 30 positions. With 15° and 30° external rotation or 30° internal rotation, the corrected stem level was not more accurate than the uncorrected stem level for any femoral position.

CONCLUSION

Use of a proportion correction minimized variability in the stem-level measurements due to sagittal angulation (up to 30°) of the femur at 0° or 15° of internal rotation.

CLINICAL SIGNIFICANCE

A proportion correction can improve the accuracy of femoral stem-level assessment on craniocaudal projection radiographs, which may aid in the detection of subsidence after cementless total hip replacement.

摘要

目的

评估尸体模型中校正比例的无水泥股骨柄水平计算的准确性。

研究设计

尸体研究。

样本人群

12 具犬股骨。

方法

在 12 具尸体股骨中,以三个柄水平植入无水泥股骨柄。在矢状面(相对于射线照相板)的 0°、10°、20°、30°和 45°角以及 0°、15°和 30°内旋和外旋的情况下,获得植入股骨的头尾投影射线照片。包括实际柄长度、射线照相测量的柄长度和射线照相测量的相对于大转子的柄水平(未校正的柄水平)的比例计算得出校正后的柄水平。比较了柄水平测量的误差。

结果

在 0°或 15°内旋时,在 30 个测试股骨位置中的 2 个位置,真实的柄水平与校正后的柄水平相差超过 0.5mm。在 10 个位置中,与未校正的柄水平相差超过 0.5mm。在 15°和 30°外旋或 30°内旋时,对于任何股骨位置,校正后的柄水平都不如未校正的柄水平准确。

结论

在 0°或 15°内旋时,由于股骨矢状面角度(最高 30°)的变化,使用比例校正可最大程度地减少柄水平测量的变异性。

临床意义

比例校正可以提高在头尾投影射线照片上评估股骨柄水平的准确性,这可能有助于检测无水泥全髋关节置换术后的下沉。