Ponce A Z, McDavid W D, Lundeen R C, Morris C R
Oral Surg Oral Med Oral Pathol. 1986 Jun;61(6):645-8. doi: 10.1016/0030-4220(86)90112-x.
A series of radiographs of a tissue-equivalent phantom was taken in order to arrive at satisfactory radiographic techniques for the use of rare earth screens with the Panorex II. The radiation reduction needed for use of the rare earth screens was made with various filters. There was a 40% to 60% reduction in patient exposure for all of the rare earth exposures. Raters agreed that all films that were ranked were diagnostically acceptable, although differences were noted.
为了获得使用稀土增感屏与全景X线机II型时令人满意的X线摄影技术,拍摄了一系列组织等效体模的X线片。使用各种滤线器实现了使用稀土增感屏所需的辐射剂量减少。所有稀土增感屏曝光时患者的辐射剂量减少了40%至60%。评估者一致认为,所有评级的胶片在诊断上都是可接受的,尽管存在差异。